Phantom-based evaluation of near-infrared intracranial hematoma detector performance
Jianting Wang1, Jonathan Lin1, Yu Chen2
1U.S. Food and Drug Administration, Center for Devices and Radiological Health, Silver Spring, Maryla, United States.
Abstract:
Near-infrared spectroscopy (NIRS) is emerging as a rapid, low-cost approach for point-of-care triage of hematomas resulting from traumatic brain injury. However, there remains a lack of standardized test methods for benchtop performance assessment of these devices and incomplete understanding of relevant light-tissue interactions. We propose a phantom-based test method for systems operating near the 800-nm oxy-/deoxy-hemoglobin isosbestic point and implement it to evaluate a clinical system. Semi-idealized phantom geometries are designed to represent epidural/subdural, subarachnoid, and intracerebral hemorrhages. Measurements of these phantoms are made with a commercial NIRS-based hematoma detector to quantify the effect of hematoma type, depth, and size, as well as measurement repeatability and detector positioning relative to the hematoma. Results indicated high sensitivity to epidural/subdural and subarachnoid hematomas. Intracerebral hematomas are detectable to a maximum depth of ∼2.5 cm, depending on thickness and diameter. The maximum lateral detection area for the single-emitter/single-collector device studied here appears elliptical and decreases strongly with inclusion depth. Overall, this study provides unique insights into hematoma detector function and indicates the utility of modular polymer tissue phantoms in performance tests for emerging NIRS-based cerebral diagnostic technology.
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II
Infrared (IR) Spectroscopy: Overview
Different compounds display unique properties due to their...
Exercise and Muscle Performance
Endurance exercises
Endurance exercises involve running, swimming, or cycling, which require repetitive movements with low force output. When a person engages in endurance exercise, a few noticeable changes occur in their skeletal muscles. For instance, the number of capillaries...