Multicenter experience with the Evolution RL mechanical sheath for lead extraction using a stepwise approach: Safety,
Federico Migliore1, Martina Testolina1, Antonio Sagone2
1Department of Cardiac, Thoracic and Vascular Sciences, University of Padova, Padova, Italy.
Background:
In addition to the Evolution RL sheath, tools by Cook Medical (Bloomington, IN, USA), supporting lead extraction (LE), are available. Data on their use are not reported in detail in previous studies. Moreover, data regarding outcome are lacking. The aim was to evaluate the safety and effectiveness of the Evolution sheath (Evolution RL and Evolution Shortie, Cook Medical) by using a stepwise approach with the available extraction tools and the outcome.
Methods:
A total of 393 leads in 198 consecutive patients were removed with the Evolution RL sheath and ancillary tools using a stepwise approach.
Results:
The main indication for LE was infection in 125 (63.1%) cases. The mean implant duration was 95.4 ± 59.7 months. According to our stepwise approach, the Evolution Shortie RL sheath was used in all cases and complete LE was achieved in 24 (12.2%) cases. The Evolution RL was used in 174 (87.8%) cases and the SteadySheath Evolution tissue stabilization sheath (Cook Medical) in 87 (44%) because of tenacious fibrosis anchored targeted leads. Compression coil (OneTie, Cook Medical) was used in 141 (71%) cases. Complete procedural success rate, clinical success rate, and lead removal with clinical success rate were 97%, 99%, and 99.5%, respectively. One major complication (0.5%) and 10 (5%) minor complications were encountered. During a mean time follow-up of 12 ± 9 months, 14 (7%) patients died. Predictors of mortality included impaired renal function (HR 5.7; 95% CI 1.9-17.6; P = 0.002), extraction because of infection (hazard ratio [HR] 4.0; 95% confidence interval [CI] 1-18.1; P = 0.045), and diabetes (HR 3.2; 95% CI 1.1-9.8; P = 0.036).
Conclusions:
Lead extraction using the Evolution RL bidirectional rotational mechanical sheath and ancillary tools in a systematic stepwise approach was effective and safe.
More Related Videos
Related Concept Videos
Extraction: Effects of pH
The Evidence for Evolution
RL Circuits
Convergent Evolution
RL Circuit without Source
Applying Kirchhoff's voltage law around the loop of the circuit and substituting the voltages across the inductor and resistor yields a first-order differential equation. A logarithmic equation is obtained by rearranging the terms in this equation,...
RL Circuit with Source
The transient response of the circuit is its temporary reaction to the sudden application of the DC source. This response is characterized by a current that exponentially decays to zero as time approaches infinity. During this transitional period, the inductor behaves like a short circuit, causing the source...