Volumetric modulated arc therapy planning based on virtual monochromatic images: Effect of inaccurate CT numbers on
Shingo Ohira1, Riho Komiyama2, Tsukasa Karino2
1Department of Radiation Oncology, Osaka International Cancer Institute, Osaka, Japan; Department of Medical Physics and Engineering, Osaka University Graduate School of Medicine, Suita, Japan.
Purpose:
Though virtual monochromatic images (VMIs) at low energy levels can improve image quality, the measured Hounsfield unit (HU) values can be inaccurate. We assessed the dosimetric error due to inaccurate HU estimation in volumetric modulated arc therapy (VMAT) planning.
Methods:
Based on the VMIs at 50 keV (VMI50keV), 77 keV (VMI77keV) and single-energy CT (SECT) image for a phantom with different sizes, lookup tables (LUTL and LUTS) were created. Using an anthropomorphic phantom (head and spine regions), VMAT plans were generated based on VMI50keV, VMI77keV and SECT using the corresponding LUTL, and then, the doses were re-calculated using LUTS. For clinical cases, 30 VMAT plans (prostate, brain, and spine cases) were generated based on VMI50keV and VMI77keV.
Results:
In the anthropomorphic phantom study, the difference in the dosimetric parameters for planning target volume (PTV) in the VMAT plan based on the VMI77keV was smallest (within 0.1 Gy) among three types of treatment planning approach. In clinical cases, in general, the differences of the 3-dimensional gamma passing rate and dosimetric parameters in the treatment plans based on the VMI50keV were larger than those in the VMI77keV. Especially for brain cases, the difference for PTV was more prominent when AXB was used (the maximum difference was 0.5 Gy) than AAA.
Conclusions:
The dosimetric error due to the inaccurate HU estimation was larger in the VMIs at low energy levels. This may be clinically insignificant, but should be avoided in the VMAT treatment planning.
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Drug Dosing in Renal Diseases: Dose Adjustments Based on Drug Clearance and Elimination Rate Constant
Virtual Work
In static equilibrium, a body can experience an imaginary or virtual movement, such as displacement or rotation. The virtual work done by a force is equal to the dot product of force and virtual displacement in the direction of the force. When it comes to virtually rotating a...
Sampling Plans
Random sampling is a method where each member of the population has an equal chance of being selected for the sample. It involves selecting individuals randomly, often using random number generators or lottery-type methods. For example, when analyzing the properties of a...