Comparison of isotope pattern deconvolution and calibration curve quantification methods for the determination of
J Pitarch-Motellón1, N Fabregat-Cabello2, C Le Goff3
1Research Institute for Pesticides and Water, Universitat Jaume I, Castelló, Spain.
Abstract:
A Liquid Chromatography coupled to tandem mass spectrometry (LC-MS/MS) based method have been developed for the determination of the main estrogen compounds -estrone (E1) and 17β-estradiol (E2)- in human serum. Two isotope dilution mass spectrometry (IDMS) quantification procedures have been used: a classical calibration curve-based method were compared to a recently developed isotope pattern deconvolution (IPD) method. IPD is based on isotopic abundance measurements and multiple linear regression. Validation was performed in terms of intra-assay repeatability (n = 5), inter-assay reproducibility (n = 9) and accuracy using spiked steroid-free serum at 5 concentration levels and 3 certified reference materials. Both methodologies meet EMEA requirements yielding recoveries between 79-106% and coefficient of variations of 1.7-8.3% along all experiments. Limits of quantification as low as 5 ng/L were achieved. 40 real samples were analysed for comparison purposes showing a great correlation between calibration and IPD concentration values. Real samples were also quantified by routine immunoassay analysis, which showed a significant proportional bias of 2.55 for E1 and good correlation for E2. While methods were considered suitable for routine or countercheck analysis within the context of hospital's needs, IPD has demonstrated to be faster and cost saving.
More Related Videos
04:17Author Spotlight: Advancing Cellular and Protein Engineering to Control Biological Functions and Develop Novel Therapies
Published on: September 27, 2024
12:35Comparison of Three Different Methods for Determining Cell Proliferation in Breast Cancer Cell Lines
Published on: September 3, 2016
Related Concept Videos
Calibration Curves: Correlation Coefficient
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...
Isotopes
An element's atomic mass, or weight,...
Deconvolution
Deconvolution involves several mathematical techniques to derive the impulse response. One common approach is polynomial division. In this method, the input and output sequences are treated as coefficients of...
Elements: Chemical Symbols and Isotopes
Some symbols are derived from the common English name of the element; others are abbreviations of the name in another language — Latin, Greek or German. For example, the symbol for aluminum (common name)...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...