Does higher field strength translate into better diagnostic accuracy? A prospective comparison of breast MRI at 3 and
Matthias Dietzel1, Evelyn Wenkel1, Matthias Hammon1
1Department of Radiology, University Hospital Erlangen, Maximiliansplatz 1, 91054, Erlangen, Germany.
Objective:
We performed a prospective monocentric study to inter-individually compare the diagnostic accuracy of breast MRI at 1.5 T and 3 T.
Methods:
During a consecutive period of 12 months all 982 patients receiving breast MRI according to standard indications (problem solving, preoperative staging) were randomized to one 3 T or 1.5 T scanner both equipped with dedicated 16-channel breast coils. Protocols at 1.5 T and 3 T were optimized and in line with international recommendations. Due to restricted time slot availabilities, the randomization-key was defined as 1/10 (3 T/1.5 T). All examinations were read by two experts in breast MRI (>25 and 8 years of experience) who assigned a BI-RADS category per breast. Histopathological verification or long-term MRI follow-up (>24 months) served as standard of reference. Results were analyzed using cross tabulations, standard estimates of diagnostic accuracy, Chi-square test and Mann-Whitney U test (alpha = 5%).
Results:
1961 breasts were included. 1746 (89%) were imaged at 1.5 T and 215 at 3 T (11%). The 1.5 T and 3 T study cohort did not show significant differences regarding patient age (P = 0.71), results of the reference standard (P = 0.09) and indication for MRI (P = 0.53). Overall Sensitivity (94.7%), Specificity (91.4%), Accuracy (91.9%) and Negative Predictive Value (99.0%) were within the range of the literature. Pairwise comparison of Sensitivity (1,5 T/3 T: 94.1/97.9%), Specificity (91.6/%89.3%), Accuracy (92.0%/91.2%) and Negative Predictive Value (98.9%/99.3%) were without significant differences (P = 0.29-0.74).
Conclusion:
In this prospective monocentric study, we identified comparably high diagnostic accuracy for both 1.5 T and 3 T breast MRI. Both 1.5 T and 3 T are equally suited for breast imaging.
Related Concept Videos
Improving Translational Accuracy
Improving Translational Accuracy
Translation
Translation is the process of synthesizing proteins from the genetic information carried by messenger RNA (mRNA). Following transcription, it constitutes the final step in the expression of genes. This process is carried out by ribosomes, complexes of protein and specialized RNA molecules. Ribosomes, transfer RNA (tRNA), and other proteins produce a chain of amino acids—the polypeptide—as the end product of translation.
Translation Produces the Building Blocks of...
Uncertainty in Measurement: Accuracy and Precision
Acid Strength and Molecular Structure
In the absence of any leveling effect, the acid strength of binary compounds of hydrogen with nonmetals (A) increases as the H-A bond strength decreases down a group in the periodic table. For group 17, the order of increasing acidity is HF < HCl < HBr < HI. Likewise, for group 16, the order of increasing acid strength is H2O < H2S < H2Se < H2Te. Across a row in the periodic table, the acid strength of binary hydrogen compounds increases with increasing...
Initiation of Translation
First, the initiator tRNA must be selected from the pool of elongator tRNAs by eukaryotic initiation factor 2 (eIF2). The initiator tRNA (Met-tRNAi) has conserved sequence elements including modified bases at...