The dosimetric effect of residual breath-hold motion in pencil beam scanned proton therapy - An experimental study
Jenny Gorgisyan1, Antony J Lomax2, Per Munck Af Rosenschold3
1Paul Scherrer Institute, ETH Domain, Switzerland; Department of Oncology, Rigshospitalet Copenhagen University Hospital, Denmark; Niels Bohr Institute, University of Copenhagen, Denmark.
Background And Purpose:
Motion management in the treatment of lung cancer is necessary to assure highest quality of the delivered radiation therapy. In this study, the breath-hold technique is experimentally investigated for pencil beam scanned (PBS) proton therapy, with respect to the dosimetric effect of residual breath-hold motion.
Material And Methods:
Three-dimensional (3D)-printed tumours extracted from CT scans of three patients were inserted into a dynamic anthropomorphic breathing phantom. The target was set up to move with the individual patient's tumour motion during breath-hold as previously assessed on fluoroscopy. Target dose was measured with radio-chromic film, and both single field uniform dose (SFUD) and intensity-modulated proton therapy (IMPT) plans were delivered. Experiments were repeated for each patient without any motion, to compute the relative dose deviation between static and breath-hold cases.
Results:
SFUD plans showed small dose deviations between static and breath-hold cases, as evidenced by the gamma pass rate (3%, 3 mm) of 85% or higher. Dose deviation was more evident for IMPT plans, with gamma pass rate reduced to 50-70%.
Conclusions:
The breath-hold technique is robust to residual intra-breath-hold motion for SFUD treatment plans, based on our experimental study. IMPT was less robust with larger detected dose deviations.
More Related Videos
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
10:03Proton Transfer and Protein Conformation Dynamics in Photosensitive Proteins by Time-resolved Step-scan Fourier-transform Infrared Spectroscopy
Published on: June 27, 2014
Related Concept Videos
Breathing
Residual Stresses
Residual Plots
When the residual values are plotted against the variable x, it is called a residual...
Serum Laboratory Studies, Stool Test, Breath Test
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Residual Stresses in Circular Shafts