A comparative in vitro study of standard facemask jet nebulization and high-flow nebulization in bronchiolitis
Juan Valencia-Ramos1, Juan Arnaez2, José Manuel Benito3
1a Paediatric Intensive Care Unit , Hospital Universitario de Burgos , Burgos , Spain.
Abstract:
Aim of Study: The use of a nebulizer paired with high-flow nasal cannulas (HFNC) has been proposed for drug delivery in bronchiolitis. Particle size nebulized is a relevant factor determining the efficacy of the nebulization. We replicated in vitro the theoretical parameters most widely used in bronchiolitis and we compared the size of the droplet nebulized with a standard nebulizer and a nebulizer integrated into HFNC. Materials and Methods: We used laser diffraction to analyze the particle size nebulized (volume median diameter Dv50). The standard system was a jet nebulizer connected to a facemask with a flow rate of 8 L/min (JN). Three designs were used as nebulizers integrated into HFNC: a vibrating mesh nebulizer set 1) before (HFNC-BH) and 2) after (HFNC-AH) the humidifier, and 3) a jet nebulizer connected before the nasal cannula (HFNC-BNC). HFNC was used with neonatal (3-8 L/min) and infant cannulas (8-15 L/min). Results: Droplet size was similar among the three drugs studied. A lower particle size was obtained when using the nebulization system integrated into HFNC compared to the standard nebulizer, regardless of the flow rate and the nasal cannula used when the position of the nebulizer was before the nasal cannula (p < 0.05): 6.89 µm (JN), 2.49 µm (HFNC-BNC 3 L/min), 2.59 µm (HFNC-BNC 5 L/min), 2.44 µm (HFNC-BNC 8 L/min), 3.22 µm (HFNC-BNC 10 L/min), 3.23 µm (HFNC-BNC 13 L/min), 3.16 µm (HFNC-BNC 15 L/min). The particle size was lower in HFNC-BF compared to the HFNC-AH using neonatal nasal cannula (3-8 L/min) (p < 0.05). Conclusion: The use of a nebulizer integrated with HFNC has shown promising results in an experimental scenario of bronchiolitis. The particle size achieved with the nebulizer placed before the humidifier is equivalent to the one obtained via conventional nebulization, and it is even smaller when the integrated nebulizer is placed before the nasal cannulas.
Related Concept Videos
Free Jet
Standard Electrode Potentials
Measurement: Standard Units
Comparing Copy Number Variations and SNPs
Copy number variations or CNVs are the structural variations that cover more than 1kb of DNA sequence. The single nucleotide polymorphism (SNP), on the other hand, is a single nucleotide change or a point mutation that is found in more than 1%...
Standard Enthalpy of Formation
Standard Entropy Change for a Reaction