Study of Nickel Levels in Patients with Atrial Septal Defect Undergoing Amplatzer Device Closure
Ganesh Narayana1, Ganavi Ramagopal2, Bhanu Duggal3
1Department of Cardiology, Dhanalakshmi Srinivasan Medical College and Hospital, Perambalur, Tamil Nadu, India.
Abstract:
Transcatheter closure of atrial septal defects (ASDs) using nickel titanium Naval Ordnance Laboratory (Nitinol)-containing devices (Amplatzer, AGA Medical, Golden Valley, Minnetosa) is being widely practiced. Controversies still exist regarding the release of nickel from these devices and the allergy and other ill effects of nickel and also the duration of antiplatelet therapy to aid endothelization of the device.
Objective:
To study the nickel levels in patients who underwent ASD closure with platinum-coated Nitinol-containing Amplatzer septal occluder.
Methodology:
A prospective study was conducted on 25 patients, between 6 years and 40 years of age who underwent ASD closure with Amplatzer septal occlude sizes ranging from 12 to 36 mm from January 2009 to January 2010. Blood nickel levels were estimated using atomic absorption photometry before and 24 h after the procedure and later at 1 month, 3 months, and 6 months postprocedure. A value of <2 mcg/dl was considered to be normal. Statistical analysis was performed by the use of the Wilcoxon test.
Results:
The blood nickel levels at mean baseline, 24 h, 1 month, 3 months, and 6 months postprocedure were 1.05, 1.39, 0.98, 0.79, and 0.74 (mcg/dl), respectively.
Conclusions:
Percutaneous ASD closure using Nitinol devices can be carried out safely without any significant ill effects related to nickel release.
More Related Videos
08:00Reduction of Iatrogenic Atrial Septal Defects with an Anterior and Inferior Transseptal Puncture Site when Operating the Cryoballoon Ablation Catheter
Published on: June 15, 2015
08:10Estimating Bilateral Atrial Function by Cardiovascular Magnetic Resonance Feature Tracking in Patients with Paroxysmal Atrial Fibrillation
Published on: July 20, 2022
Related Concept Videos
Lumber Defects
Shakes are minor fractures that run along or across the wood's annual rings, while wane is...
Angle Closure Glaucoma: Treatment
High-Level and Low-Level Awareness
Leveling Effect
Bioavailability Study Design: Healthy Subjects Versus Patients
Levels of Organization
Molecules Are Composed of Atoms, and Biomolecules Are Assembled from Molecules:
The most basic levels include atoms, molecules, and biomolecules. Atoms, the smallest unit of ordinary matter, are composed of a nucleus and electrons. Molecules...