Two-dimensional ultrasound-computed tomography image registration for monitoring percutaneous hepatic intervention
Robert M Pohlman1, Michael R Turney1, Po-Hung Wu2
1Department of Medical Physics, University of Wisconsin School of Medicine and Public Health, University of Wisconsin-Madison, Madison, WI, 53706, USA.
Purpose:
Deformable registration of ultrasound (US) and contrast enhanced computed tomography (CECT) images are essential for quantitative comparison of ablation boundaries and dimensions determined using these modalities. This comparison is essential as stiffness-based imaging using US has become popular and offers a nonionizing and cost-effective imaging modality for monitoring minimally invasive microwave ablation procedures. A sensible manual registration method is presented that performs the required CT-US image registration.
Methods:
The two-dimensional (2D) virtual CT image plane that corresponds to the clinical US B-mode was obtained by "virtually slicing" the 3D CT volume along the plane containing non-anatomical landmarks, namely points along the microwave ablation antenna. The initial slice plane was generated using the vector acquired by rotating the normal vector of the transverse (i.e., xz) plane along the angle subtended by the antenna. This plane was then further rotated along the ablation antenna and shifted along with the direction of normal vector to obtain similar anatomical structures, such as the liver surface and vasculature that is visualized on both the CT virtual slice and US B-mode images on 20 patients. Finally, an affine transformation was estimated using anatomic and non-anatomic landmarks to account for distortion between the colocated CT virtual slice and US B-mode image resulting in a final registered CT virtual slice. Registration accuracy was measured by estimating the Euclidean distance between corresponding registered points on CT and US B-mode images.
Results:
Mean and SD of the affine transformed registration error was 1.85 ± 2.14 (mm), computed from 20 coregistered data sets.
Conclusions:
Our results demonstrate the ability to obtain 2D virtual CT slices that are registered to clinical US B-mode images. The use of both anatomical and non-anatomical landmarks result in accurate registration useful for validating ablative margins and comparison to electrode displacement elastography based images.
More Related Videos
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
05:39Three-Dimensional Printing Guide Template Assisted Percutaneous Vertebroplasty PVP
Published on: October 17, 2019
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Nursing Interventions I: Taxonomy of Nursing Interventions
A nursing intervention is a treatment or action based on scientific concepts and knowledge from the nursing, behavioral, and physical sciences. Identifying and prioritizing nursing interventions based on the desired outcome...
Nursing Interventions II: Selecting and Classifying the Nursing Interventions
Community Based Intervention
Foundations of Community Mental Health Programs
Central to the success of community-based interventions is the...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET