Measurement of urinary free cortisol by LC-MS-MS: adoption of a literature reference range and comparison with our
L Bianchi1, B Campi2, M R Sessa1
1Laboratory of Chemistry and Endocrinology, University Hospital of Pisa, Via Paradisa 2, 56124, Pisa, Italy.
Purpose:
One of the best indicators of adrenal gland dysfunction is the level of free cortisol measured in the 24-h urine (UFC) which faithfully reflects the level of biologically active serum cortisol not subjected to circadian variations. Liquid chromatography coupled with tandem mass spectrometry (LC-MS-MS) is a sensitive, accurate and precise method recently available in routine laboratories that could remedy interference problems of immunoassays.
Methods:
In this study, a literature reference range for UFC measured by LC-MS-MS was verified, and UFC values measured by LC-MS-MS and immunoassay were compared. Immunometric UFC measurement was performed by ACCESS CORTISOL assay without preliminary extraction, using Beckman Coulter UniCel DxI 600 highly automated platform. Liquid chromatography-tandem mass spectrometry UFC measurement was performed by a home-made validated method using cortisol-D4 as internal standard with preliminary deproteinization of urinary samples by centrifugal filter and injection on reverse-phase column. Cortisol was analyzed in positive ion mode with an ESI interface.
Results:
The reference interval from literature (11-70 μg/day) was confirmed by results obtained for healthy study group. Comparison study of the two methods highlighted a constant and proportional systematic error with a general tendency to overestimate results for the in-use method.
Conclusions:
In conclusion, the direct immunometric method overestimates UFC results with respect to liquid chromatography-tandem mass spectrometry which represents the reference method. The literature reference range 11-70 μg/day was confirmed and can be adopted by our lab that will shift all UFC tests performed in routine to the mass spectrometry-based method, satisfying clinicians' request.
More Related Videos
Related Concept Videos
Range
15.9; 16.1; 15.2; 14.8; 15.8; 15.9; 16.0; 15.5
Measurements of the amount of soda in a 16-ounce can vary since different subjects record these measurements or since the exact amount - 16 ounces of liquid, was not...
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Variation: Normal Distribution, Range, and Standard Deviation
Renal Drug Clearance: Comparison Between Renal Excretion Methods
Renal clearance is often associated with the renal glomerular filtration rate (GFR), which represents the rate at which plasma is filtered through the glomeruli in the kidney. When drug reabsorption is minimal and there is no active secretion, renal clearance is closely related to the...
Inertial Frames of Reference
Non-inertial Frames of Reference