Pacemaker Dysfunction due to a Large Thrombus on Ventricular Lead
Zhen Xia1, Jianhua Yu1, Juxiang Li1
1Department of Cardiology, The Second Affiliated Hospital of Nanchang University, Nanchang, China.
Background:
Pacemaker lead-related thrombosis is a rare but severe complication in patients with pacing lead implantation in the right ventricle. We present a case with recurrent syncope after single-chamber implantable cardioverter defibrillator (ICD) implantation. Pacing lead-related thrombosis was observed during open-heart surgery. This induced intermittent pacemaker dysfunction and recurrent syncope.
Case Presentation:
A 67-year-old male patient presented with frequent episodes of syncope and a history of dilated cardiomyopathy and paroxysmal ventricular tachycardia. Normal coronary angiography was found, and therefore a single-chamber ICD was implanted into the right ventricle to prevent cardiac events in 2013. However, he was referred to our hospital because of recurrent syncope 3 to 4 years after ICD implantation. A comprehensive investigation was performed to find out the etiology for the recurrent syncope. Pacing lead thrombosis was finally observed during open-heart surgery, which can introduce intermittent pacemaker dysfunction. After the thrombus was removed and the lead was separated from the posterior leaflet of the tricuspid valve, the ICD functioned normally after reprogramming. Oral anticoagulant was prescribed after discharging. During the 1-year follow-up period, this patient was free of syncope.
Conclusions:
This case illustrated that pacemaker lead-associated thrombosis should be considered when the cardiac implantable electronic device fails to prevent patients from having cardiac events. Oral anticoagulant might be important for preventing thrombosis among patients with ICD implantation into the right ventricle.
Related Concept Videos
Phase-lead and Phase-lag Controllers
Time and frequency -Domain Interpretation of Phase-lead Control
The design of phase-lead control involves the strategic placement of poles and zeros to balance steady-state error and system...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Anatomy of the Heart
Factors Affecting Solubility
Precipitation of Ions
The equation that describes the equilibrium between solid calcium carbonate and its solvated ions is: