Method to assess the mismatch between the measured and nominal parameters of transcranial magnetic stimulation
Leonardo Rakauskas Zacharias1, André Salles Cunha Peres2, Victor Hugo Souza3
1Departamento de Física, Faculdade de Filosofia Ciências e Letras de Ribeirão Preto, Universidade de São Paulo, Ribeirão Preto, SP, Brazil; Departamento de Neurociências e Ciências do Comportamento, Faculdade de Medicina de Ribeirão Preto, Universidade de São Paulo, Ribeirão Preto, SP, Brazil.
Background:
Small variations in TMS parameters, such as pulse frequency and amplitude may elicit distinct neurophysiological responses. Assessing the mismatch between nominal and experimental parameters of TMS stimulators is essential for safe application and comparisons of results across studies.
New Method:
A search coil was used to assess exactness and precision errors of amplitude and timing parameters such as interstimulus interval, the period of pulse repetition, and intertrain interval of TMS devices. The method was validated using simulated pulses and applied to six commercial stimulators in single-pulse (spTMS), paired-pulse (ppTMS), and repetitive (rTMS) protocols, working at several combinations of intensities and frequencies.
Results:
In a simulated signal, the maximum exactness error was 1.7% for spTMS and the maximum precision error 1.9% for ppTMS. Three out of six TMS commercial devices showed exactness and precision errors in spTMS amplitude higher than 5%. Moreover, two devices showed amplitude exactness errors higher than 5% in rTMS with parameters suggested by the manufactures.
Comparison With Existing Methods:
Currently available tools allow characterization of induced electric field intensity and focality, and pulse waveforms of a single TMS pulse. Our method assesses the mismatch between nominal and experimental values in spTMS, ppTMS and rTMS protocols through the exactness and precision errors of amplitude and timing parameters.
Conclusion:
This study highlights the importance of evaluating the physical characteristics of TMS devices and protocols, and provides a method for on-site quality assessment of multiple stimulation protocols in clinical and research environments.
More Related Videos
12:19Utilizing Transcranial Magnetic Stimulation to Study the Human Neuromuscular System
Published on: January 20, 2012
13:56The Use of Magnetic Resonance Spectroscopy as a Tool for the Measurement of Bi-hemispheric Transcranial Electric Stimulation Effects on Primary Motor Cortex Metabolism
Published on: November 19, 2014
Related Concept Videos
Mismatch Repair
Mismatch Repair
The Mutator Protein Family Plays a Key Role in DNA Mismatch Repair
The human genome has more than 3 billion base pairs of DNA per cell. Prior to cell division, that vast amount of genetic...
Nominal Level of Measurement
The data that cannot be measured but can be grouped into categories fall under the nominal level of measurement. Data that is measured using a nominal...
Transformers with Off-Nominal Turns Ratios
Sign Test for Nominal Data
For example, consider a...
Wave Parameters