Quantitative evaluation of 4D Cone beam CT scans with reduced scan time in lung cancer patients
Abigail Bryce-Atkinson1, Thomas Marchant2, John Rodgers3
1Division of Cancer Sciences, School of Medical Sciences, Faculty of Biology, Medicine and Health, The University of Manchester, UK.
Purpose:
Image guided radiotherapy (IGRT) based on respiration correlated cone-beam CT (4D-CBCT) provides accurate tumour localisation in lung cancer patients by taking into account respiratory motion when deriving setup correction. However, 4D-CBCT scan times are typically longer than for acquisition of 3D-CBCT scans, e.g. 4 min. This work aims to quantitatively evaluate the effect of reduced scan times on 4D-CBCT image quality and registration accuracy in lung cancer patients.
Methods And Materials:
Scan times down to 1 min were simulated by retaining only projection images corresponding to every second, third or fourth respiratory cycle in forty-four 4D-CBCTs from 15 lung cancer patients. In addition twenty 2-minute scans were acquired for 12 lung cancer patients. Image quality was quantified by assessing registration accuracy in the shorter scan times, comparing to the 4-minute scan registration result where available as reference.
Results:
Use of 2-minute scans had little impact on registration accuracy or ability to detect tumour motion: automatic registration accuracy was within 2 mm in 6/8 scans analysed with 2-minute acquisitions, and 96.6% of registration discrepancies were within 2 mm for the simulated scans. When the scan time simulated was below 2 min, automatic registration results still agreed within 2 mm for 84.7% of scans, however visual image quality was considerably degraded.
Conclusion:
A 4D-CBCT acquisition time of 2 min produces scans of sufficient image quality for IGRT in most lung cancer patients, as demonstrated quantitatively by assessing the impact on automatic registration accuracy in simulated and real acquisitions.
More Related Videos
06:51Utilizing 18F-FDG PET/CT Imaging and Quantitative Histology to Measure Dynamic Changes in the Glucose Metabolism in Mouse Models of Lung Cancer
Published on: July 21, 2018
06:55Scanning Light Scattering Profiler SLPS Based Methodology to Quantitatively Evaluate Forward and Backward Light Scattering from Intraocular Lenses
Published on: June 6, 2017
Related Concept Videos
Leaky Scanning
Voltammetric Techniques: Linear-Scan (E vs Time)
Scanning Electron Microscopy
Fundamental Principles
Accelerated...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...