Prevention of Lung Bacterial Colonization With a Leak-Proof Endotracheal Tube Cuff: An Experimental Animal Study
Emanuele Rezoagli1,2,3,4, Massimo Cressoni1, Giacomo Bellani1,5
1School of Medicine and Surgery, University of Milano-Bicocca, Monza, Italy.
Background:
Endotracheal tubes with standard polyvinyl chloride cuffs create folds on inflation into the trachea, which lead to potential leakage of subglottic secretions into the lower airways and cause lung colonization and pneumonia. The use of a double-layer prototype leak-proof cuff has shown effective prevention of the fluid leakage across the cuff. We hypothesized that the use of such a leak-proof cuff could prevent lung bacterial colonization in vivo.
Methods:
To simulate patients in the ICU, 13 pigs were placed in the semirecumbent position, intubated, and mechanically ventilated for 72 h. Five animals were prospectively intubated with an endotracheal tube with a leak-proof cuff (leak-proof cuff group). Data from 8 animals previously intubated with an endotracheal tube with a standard polyvinyl chloride cuff (standard cuff group) were retrospectively analyzed. Leakage of tracheal secretions across the leak-proof cuff was tested by the macroscopic methylene blue evaluation. Arterial blood gas exchanges and microbiology were tested in all the pigs at necropsy.
Results:
In the standard cuff group, all the pigs showed heavy bacterial colonization of the lungs after 72 h of mechanical ventilation, with an overall proportion of colonized lung lobes of 92% (44/48 lobes, 8/8 animals) compared with 27% (8/30 lobes, 5/5 animals) in the leak-proof cuff group (P < .001). These results were strengthened by the absence of methylene blue in the tracheal secretions below the leak-proof cuff. Furthermore, no hypoxemia was demonstrated in the pigs in the leak-proof cuff group after the 72-h experiment (PaO /FIO change from baseline, leak-proof cuff group vs standard cuff group; median difference 332, 95% CI 41-389 mm Hg; P = .030).
Conclusions:
A new leak-proof cuff for endotracheal intubation prevented macroscopic leakage of subglottic secretions along the airways. This mechanism led to the reduction of lung bacterial colonization, which could contribute to the prevention of hypoxemia in the pigs on mechanical ventilation while in the semirecumbent position.
Related Concept Videos
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Social Proof
The Colonization of Land
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Acute Kidney Injury IV: Diagnostic Studies and Prevention