On probing intravoxel incoherent motion in the heart-spin-echo versus stimulated-echo DWI
Georg R Spinner1, Christian T Stoeck1, Linda Mathez1
1Institute for Biomedical Engineering, University and ETH Zurich, Zurich, Switzerland.
Purpose:
Mapping intravoxel incoherent motion (IVIM) in the heart remains challenging despite advances in cardiac DWI and DTI. In the present work, simulations and experimental imaging are used to compare the IVIM encoding efficiency of spin-echo- and stimulated-echo-based DWI/DTI for assessing myocardial perfusion.
Methods:
Using normalized phase distributions and statistical models of capillary networks derived from histological studies, along with typical diffusion gradient waveforms for in vivo cardiac DWI/DTI, Monte Carlo simulations were performed. The simulation results were compared to IVIM measurements of perfused porcine hearts regarding both magnitude and phase modulation. An IVIM tensor model was used to account for anisotropy of the capillary network, and potential bias of parameter estimation was reported based on simulations.
Results:
Both computer simulations and experimental data demonstrate a low sensitivity of spin-echo DWI/DTI sequences for IVIM parameters, whereas stimulated-echo-based DWI/DTI with typical mixing times can differentiate between no-flow baseline and perfused myocardium (+129% IVIM-derived flow). In addition, ischemic territories induced by coronary occlusion could be successfully detected. With increasing order of motion compensation (M0/M1/M2) of the diffusion encoding gradients, as required for cardiac in vivo spin-echo DWI/DTI, the low IVIM sensitivity of spin-echo DWI/DTI decreased further in simulations: maximum attenuations of perfusion compartment 52/13/5% (b = 500 s/mm2 ).
Conclusion:
Given the short encoding time of spin-echo-based DWI/DTI sequences, a limited perfusion sensitivity results, in particular in combination with motion-compensated diffusion gradients. In contrast, stimulated-echo based DWI/DTI has the potential to identify perfusion changes in cardiac IVIM in vivo.
More Related Videos
Related Concept Videos
Echo
Imagine the sound is reflected back to the ears. Assuming that the source is very close to the human, the difference between hearing the two sounds—the emitted sound and the reflected sound—may be more than the minimum time for perceiving distinct sounds. If this is the case,...
Spin–Spin Coupling Constant: Overview
Qualitatively, any spin plus-half nucleus polarizes the spins of its electrons to the minus-half state. Consequently, the paired electron in the hydrogen–carbon bond must...
NMR Spectroscopy: Spin–Spin Coupling
Spin–Spin Coupling: One-Bond Coupling
Spin–Spin Coupling: Two-Bond Coupling (Geminal Coupling)
The central atom need not be NMR-active because its electrons are affected by the electron polarization of the spin-active atoms. However, spin information is transmitted less effectively than in one-bond coupling, and 2J values are usually weaker than 1J values. The energy of...
Spin–Spin Coupling: Three-Bond Coupling (Vicinal Coupling)
The extent of coupling depends on the C‑C bond length, the two H‑C‑C angles, any electron-withdrawing substituents, and the dihedral angle between the involved orbitals. The...