Feasibility of Micro-Computed Tomography Imaging for Direct Assessment of Surgical Resection Margins During
Natasja N Y Janssen1, Maartje van Seijen2, Claudette E Loo3
1Department of Radiation Oncology, Netherlands Cancer Institute, Amsterdam, the Netherlands.
Background:
To analyze the feasibility and accuracy of micro-computed tomography (micro-CT) for surgical margin assessment in breast excision specimen.
Materials And Methods:
Two data sets of 30 micro-CT scans were retrospectively evaluated for positive resection margins by four observers in two phases, using pathology as a gold standard. Results of phase 1 were evaluated to define micro-CT evaluation guidelines for phase 2. Interobserver agreement was also assessed (kappa). In addition, a prospective study was conducted in which 40 micro-CT scans were directly acquired, reconstructed, and evaluated for positive resection margins by one observer. A suspect positive resection margin on micro-CT was annotated onto the specimen with ink, enabling local validation by pathology. Main outcome measures were accuracy, sensitivity, specificity, and positive predictive value (PPV).
Results:
Average accuracy, sensitivity, specificity, and PPV for the four observers were 63%, 38%, 70%, and 22%, respectively, in phase 1 and 72%, 40%, 78%, and 26%, respectively, in phase 2. The interobserver agreement was fair [kappa (range), 0.31 (0.12-0.80) in phase 1 and 0.23 (0-0.43) in phase 2]. In the prospective study 70% of the surgical resection margins were correctly evaluated. Ten specimens were annotated for positive resection margins, which correlated with three positive and three close (<1 mm) margins on pathology. Sensitivity, specificity, and PPV were 38%, 78%, and 30%, respectively.
Conclusions:
Micro-CT imaging of breast excision specimen has moderate accuracy and considerable interobserver variation for analysis of surgical resection margins. Especially sensitivity and PPV need to be improved before micro-CT-based margin assessment can be introduced in clinical practice.
More Related Videos
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
What is Conservation Biology?
Conservation of Small Populations
Conserved Binding Sites
Binding sites are often located in large pockets, and if their location on a protein’s surface is unknown, it can be predicted using various approaches. The energetic method computationally...
Margin of Error