A motor-driven adjustable prosthetic socket operated using a mobile phone app: A technical note
Joan E Sanders1, Joseph L Garbini2, Jake B McLean1
1Department of Bioengineering, University of Washington, Seattle, WA 98195, United States.
Medical Engineering & Physics
|April 28, 2019
Summary
This study presents an ambulatory adjustable socket system for individuals with trans-tibial limb loss. The device allows users to adjust prosthetic socket fit during walking, improving comfort and management of limb volume changes.
Area of Science:
- Biomedical Engineering
- Prosthetics and Orthotics
- Rehabilitation Technology
Background:
- Limb volume fluctuations in trans-tibial amputees impact prosthetic fit and user comfort.
- Current prosthetic sockets offer limited adjustability during ambulation, hindering effective limb volume management.
Purpose of the Study:
- To develop and evaluate a novel ambulatory adjustable socket system for trans-tibial prosthesis users.
- To enable incremental socket size adjustments during ambulation to improve prosthetic fit and user experience.
Main Methods:
- A platform system with a motor-driven cable mechanism was designed for radial socket panel adjustment.
- A mobile phone interface controlled the motor, with a proportional-integral-derivative controller managing adjustments.
- Bench testing, treadmill trials with 16 participants, and field testing with 13 participants were conducted.
Main Results:
- Bench tests demonstrated a maximum steady-state error of 0.036 mm under load.
- Treadmill testing revealed clinically acceptable fit variations between 24 mm and 114 mm in cable length.
- Field testing showed 11 of 13 participants could comfortably adjust socket size while walking.
Conclusions:
- The developed system successfully achieves incremental socket size adjustments suitable for research and development.
- Ambulatory adjustable sockets show promise for enhancing prosthetic fit and user satisfaction in trans-tibial amputees.
Related Concept Videos
Adjusting a Traverse
378
In the site survey of a four-sided traverse, internal angles are essential to ensure geometric accuracy. The survey revealed that the sum of the measured internal angles was 359 degrees and 48 minutes, which is 12 minutes less than the expected 360 degrees. This discrepancy signals an error likely arising from measurement inaccuracies during the fieldwork.To rectify this error, the adjustment process involved distributing the 12-minute shortfall equally across the four internal angles. By...
378
The Anchoring-and-Adjustment Heuristic
7.7K
In order to make good decisions, we use our knowledge and our reasoning. Often, this knowledge and reasoning is sound and solid. However, sometimes, we are swayed by biases or by others manipulating a situation. For example, let’s say you and three friends wanted to rent a house and had a combined target budget of $1,600. The realtor shows you only very run-down houses for $1,600 and then shows you a very nice house for $2,000. Might you ask each person to pay more in rent to get the...
7.7K
Renal Failure: Dose Adjustments
451
In patients with renal impairment, drugs undergo significant changes in their pharmacokinetics, which require dosage adjustments to ensure safe and effective therapy.
Reduced renal clearance and elimination rate are common outcomes of renal impairment. These alterations lead to a prolonged elimination half-life and an altered apparent volume of distribution for drugs. As a result, dosage adjustments are typically necessary to maintain optimal drug levels in the body.
However, dosage adjustments...
Reduced renal clearance and elimination rate are common outcomes of renal impairment. These alterations lead to a prolonged elimination half-life and an altered apparent volume of distribution for drugs. As a result, dosage adjustments are typically necessary to maintain optimal drug levels in the body.
However, dosage adjustments...
451
Operational Amplifiers
1.9K
The operational amplifier, often referred to as an op-amp, is a multifaceted building block of a circuit. This electronic component functions like a voltage-controlled voltage source and can also be used to create a voltage- or current-controlled current source. The design of an operational amplifier enables it to execute mathematical operations when external components like resistors and capacitors are linked to its terminals. An op-amp has the capacity to sum signals, amplify a signal,...
1.9K
Vector Operations
2.2K
Vectors are physical quantities that have both magnitude and direction. The vector operations include addition, subtraction, and scalar multiplication.
A vector multiplied by a scalar value is called scalar multiplication. The result obtained is a new vector with a different magnitude. If the scalar is positive, the direction of the vector remains the same, but if it is negative, the direction of the vector is reversed. For example, the product of the mass and velocity yields the momentum.
A vector multiplied by a scalar value is called scalar multiplication. The result obtained is a new vector with a different magnitude. If the scalar is positive, the direction of the vector remains the same, but if it is negative, the direction of the vector is reversed. For example, the product of the mass and velocity yields the momentum.
2.2K
Operant Conditioning
2.8K
Operant conditioning, a key concept in behavioral psychology, involves using reinforcement and punishment to alter the likelihood of a behavior being repeated. B.F. introduced this type of conditioning. Skinner focused on voluntary behaviors and the consequences that follow them, influencing whether these behaviors will be strengthened or diminished.
Reinforcement in operant conditioning can be positive or negative, both of which serve to increase the likelihood of a behavior. Positive...
Reinforcement in operant conditioning can be positive or negative, both of which serve to increase the likelihood of a behavior. Positive...
2.8K