Corneal higher-order aberrations measurements: Comparison between Scheimpflug and dual Scheimpflug-Placido technology
Andre L Piccinini1, Oren Golan2, Emilio A Torres-Netto3
1Keck School of Medicine of the University of Southern California, Los Angeles, California, USA; Sadalla Amin Ghanem Eye Hospital, Joinville, SC, Brazil.
Purpose:
To compare the difference and agreement of corneal higher-order aberrations (HOAs) in keratoconic eyes using Scheimpflug and dual Scheimpflug-Placido imaging systems.
Setting:
Emory University, Atlanta, Georgia, and the University of Southern California, Los Angeles, California, USA.
Design:
Retrospective between-devices reliability and agreement study.
Methods:
Patients diagnosed with keratoconus were evaluated sequentially by Scheimpflug and dual Scheimpflug-Placido devices. Differences, correlations, and agreement between values for total root mean square (RMS), trefoil, coma, and spherical aberration were analyzed, and Bland-Altman plots were generated.
Results:
Fifty eyes from 50 patients (31 men, 19 women) were evaluated. Trefoil at 30 degrees, spherical aberration, and total RMS were significantly different between groups (P < .05), whereas trefoil at 0 degrees and total coma values were not statistically different. There was a weak positive correlation between devices for trefoil at 0 degrees (r = 0.228), and a moderate positive correlation for trefoil at 30 degrees (r = 0.473), horizontal coma (r = 0.430), and for total corneal RMS (r = 0.637). Vertical coma (r = 0.816) and spherical aberration (r = 0.874) showed a strong positive correlation. The 95% limits of agreement (LoA) for absolute values were 1.963 μm for trefoil at 30 degrees, 2.449 μm for trefoil at 0 degrees, 3.530 μm for horizontal coma, 2.145 μm for vertical coma, 1.242 μm for spherical aberration, and 10.527 μm for RMS.
Conclusion:
Significant differences were found between measurements of corneal HOAs generated by Scheimpflug and dual Scheimpflug-Placido devices in patients with keratoconus, with generally limited correlations and wide LoA. HOAs measurements from these devices should not be considered equivalent.
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Hybridoma Technology
Hybridoma Selection
Commonly used fusion techniques — electroporation,...
Evolutionary Relationships through Genome Comparisons
Comparison Between Electrical And Gravitational Forces
Since both are inverse square law forces, the distance gets canceled when the ratio of the two forces is considered. Instead, the ratio of the electrical and gravitational forces depends on...
Comparison between RL and RC circuits