Related Experiment Video
Updated: Jan 25, 2026

C-arm-Free Simultaneous OLIF51 and Percutaneous Pedicle Screw Fixation in a Single Lateral Position
Published on: September 16, 2022
A new risk to the axillary nerve during percutaneous proximal humeral plate fixation using the Synthes PHILOS aiming
Khang H Dang1, Samuel S Ornell1, Guy Reyes1
1Department of Orthopaedics, University of Texas Health San Antonio, San Antonio, TX, USA.
Background:
Percutaneous aiming arms have been developed to minimize injury during placement of submuscular proximal humerus plates. The purpose of this study was to determine the risk of axillary nerve injury during percutaneous proximal humeral plate fixation using the Synthes PHILOS aiming system.
Methods:
By use of 10 fresh-frozen cadavers (20 shoulders), a 3.5-mm locking compression proximal humeral plate was fixated percutaneously to the humerus through a lateral deltoid-splitting approach using the PHILOS aiming guide. Dissection of the axillary nerve was then carried out, and measurements of its relation to the screw holes in row A through row G of the plate were taken. The lateral acromion-to-axillary nerve distance was also measured.
Results:
The axillary nerve traversed row D in every shoulder, whereas it crossed over row C in 11 shoulders and both holes in row E in 16 shoulders. The closest distance to the axillary nerve achieved was 4.5 mm, corresponding to the distal (left) screw in row B. A significant negative correlation was found for the distance from the nerve to the closest proximal and distal screws (row B and row G, respectively) in both right shoulders (ρ = -0.797; 95% confidence interval, -0.916 to -0.548) and left shoulders (ρ = -0.615; 95% confidence interval, -0.831 to -0.237).
Conclusion:
The axillary nerve traverses rows C, D, and E of the proximal humeral plate using the PHILOS aiming system. Importantly, our study is the first to demonstrate that the axillary nerve crosses over row C. Left-sided plate screws also came in closer proximity to the axillary nerve than right-sided plate screws.
Related Concept Videos
Aims Of Nursing
Fixation and Sectioning
The simplest type of preparation is the wet mount, in which the specimen is placed in a drop of liquid on the slide. A liquid specimen can be directly deposited on the slide using a dropper. Solid specimens, such as skin scraping, can be placed on the slide before adding a drop of liquid to prepare the wet mount. Sometimes the liquid is simply water, but stains are often added...
Relative Risk
Carbon-dioxide Fixation
Factors Affecting the Risk of Infection
The integrity and count of the white blood cells help the body resist pathogens and fight infection. When impaired, it reduces the body's resistance to pathogens. The acidic pH levels of the gastrointestinal, genitourinary tracts, and skin...
Factors Influencing Attraction I: Proximity