Ultrasound Tissue Characterization of Human Achilles Tendon by Stability Quantification of Echo Patterns
Published on: September 5, 2025
Inter- and intra-rater reliability of ultrasound tissue characterization (UTC) in patellar tendons
Mathijs van Ark1,2, Lucas Maciel Rabello3, Dirk Hoevenaars3
1Department of Physiotherapy, Hanze University of Applied Sciences Groningen, Groningen, The Netherlands.
Purpose:
Ultrasound tissue characterization (UTC) is used in research and clinical practice to quantify tendon structure of the patellar tendon. This is the first study to investigate the inter- and intra-rater reliability for UTC of the patellar tendon on a large scale.
Method:
Fifty participants (25 patellar tendinopathy, 25 asymptomatic) were recruited. The affected patellar tendons in symptomatic and right tendons in asymptomatic participants were scanned with UTC twice by one researcher and once by another. The same was done for contour marking (needed to analyze a UTC scan) of the tendon. Intraclass correlation coefficient (ICC (2,1)) for echo-types I, II, III, IV, aligned fibrillar structure (echo-types I + II), and disorganized structure (echo-types III + IV) were calculated. This was done for UTC scans as well as solely marking contours.
Results:
Inter-rater reliability showed fair to good ICC values for echo-types I (0.65) and II (0.46) and excellent ICC values for echo-type III (0.81), echo-type IV (0.83), aligned fibrillar structure (0.82), and disorganized structure (0.82). Intra-rater reliability showed excellent ICC values for echo-types I (0.76), III (0.88), IV (0.85), aligned fibrillar structure (0.88), and disorganized fibrillar structure (0.88) and a fair to good value for echo-type II (0.61). Contour marking showed excellent ICC values for all echo-types.
Conclusion:
This study showed that UTC scans for patellar tendons have overall good intra-rater and inter-rater reliability. To optimize reliability of UTC scans of the patellar tendon, using the same rater and using aligned fibrillar structure (echo-types I + II combined) and disorganized structure (echo-types III + IV combined) as outcome measures can be considered.
More Related Videos
08:40Isokinetic Robotic Device to Improve Test-Retest and Inter-Rater Reliability for Stretch Reflex Measurements in Stroke Patients with Spasticity
Published on: June 12, 2019
09:31Evaluation of Stem Cell Therapies in a Bilateral Patellar Tendon Injury Model in Rats
Published on: March 30, 2018
Related Concept Videos
Reliability and Validity
Distribution Reliability and Automation
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...
Naturalistic Observations
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: