Energy Window and Contrast Optimization for Single-photon Emission Computed Tomography Bremsstrahlung Imaging with
Youssef Bouzekraoui1, Farida Bentayeb1, Hicham Asmi1
1Department of Physics, LPHE, Modeling and Simulations, Faculty of Science, Mohammed V University, Rabat, Morocco.
Purpose:
In yttrium-90 (Y-90) single-photon emission computed tomography (SPECT) imaging, the choice of the acquisition energy window is not trivial, due to the continuous and broad energy distribution of the bremsstrahlung photons. In this work, we investigate the effects of the energy windows on the image contrast to noise ratio (CNR), in order to select the optimal energy window for Y-90 imaging.
Materials And Methods:
We used the Monte Carlo SIMIND code to simulate the Jaszczak phantom which consists of the six hot spheres filled with Y-90 and ranging from 9.5 to 31.8 mm in diameter. Siemens Symbia gamma camera fitted with a high-energy collimator was simulated. To evaluate the effect of the energy windows on the image contrast, five narrow and large energy windows were assessed.
Results:
The optimal energy window obtained for Y-90 bremsstrahlung SPECT imaging was 120-150 keV. Furthermore, the results obtained for CNR indicate that the high detection is only for the three large spheres.
Conclusion:
The optimization of energy window in Y-90 bremsstrahlung has the potential to improve the image quality.
More Related Videos
05:32Multimodal Bioluminescent and Positronic-emission Tomography/Computational Tomography Imaging of Multiple Myeloma Bone Marrow Xenografts in NOG Mice
Published on: January 7, 2019
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Emission Spectra
Light as Energy
Photons
A photon is a discrete electromagnetic particle or bundle of energy. Photons are characterized by their frequency, wavelength, and amplitude, similar to the properties of a wave. Waves with higher frequencies transmit more energy and have shorter wavelengths than longer wavelengths that transmit...