Background:
Current anesthesia workspaces consist of several technical devices, such as patient monitors, anesthesia machines, among others. Commonly, they are produced by different manufacturers; thus, they differ in terms of their modus operandi, user interface, and representation of alarms. Merging the information from these devices using a single joint protocol and displaying it in a single graphical user interface could lead to a general improvement in perioperative management. For this purpose, the recently approved and published Institute of Electrical and Electronics Engineers 11073 service-oriented device connectivity standard was implemented.
Objective:
This paper aims to develop and then evaluate an anesthesia workstation (ANWS) functional model in terms of usability, fulfillment of clinical requirements, and expected improvements in patient safety.
Methods:
To compare the self-developed ANWS with the conventional system, a pilot observational study was conducted at the University Hospital Aachen, Germany. A total of 5 anesthesiologists were asked to perform different tasks using the ANWS and then the conventional setup. For evaluation purposes, response times were measured and an interaction-centered usability test with an eye-tracking system was carried out. Finally, the subjects were asked to fill in a questionnaire in order to measure user satisfaction.
Results:
Response times were significantly higher when using the ANWS, but decreased considerably after one repetition. Furthermore, usability was rated as excellent (≥95) according to the System Usability Scale score, and the majority of clinical requirements were met.
Conclusions:
In general, the results were highly encouraging, considering that the ANWS was only a functional model, as well as the lack of training of the participants. However, further studies are necessary to improve the universal user interface and the interplay of the various networked devices.
Related Concept Videos
Communication
Communication
Introduction to Test of Independence
The test statistic for a test of independence is similar to that of a goodness-of-fit test:
Hypothesis Test for Test of Independence
H0: The two variables (factors)...
Role of Communication in the Nursing Process III: Evaluation and Documentation
Analysis Methods of Pharmacokinetic Data: Model and Model-Independent Approaches
The model approach uses mathematical models to describe changes in drug concentration over time. Pharmacokinetic models help characterize drug behavior in patients, predict drug concentration in the body fluids, calculate optimum dosage regimens, and evaluate the risk of toxicity. However, ensuring that the model fits the experimental data accurately...