Comparison of Preoperative Parameter Measurements Using an Optical Biometer, Automated Keratometer, and a
Ekjyot S Gill1, Christopher Le, Jacob Joseph
1Department of Surgery (W.M.M.), Baltimore Veteran Affairs Medical Center, Baltimore, MD; Department of Ophthalmology and Visual Sciences (E.S.G., C.L., W.M.M.), University of Maryland School of Medicine, Baltimore, MD; and Department of Orthopedic Surgery (J.J.), Johns Hopkins Medical Center, Baltimore, MD.
Objectives:
To assess the agreement of corneal power, corneal astigmatism, axis location, and astigmatic vector component measurements using a Lenstar LS900, a T-cone add-on, and an ARK 1S autorefractor, furthermore, to assess instrument agreement in a subset of astigmatic eyes.
Methods:
Retrospective analysis of keratometric readings from 3 instruments for 66 eyes undergoing cataract surgery at the Maryland Veterans Affairs Medical Center from November 2014 to 2015. Agreement in corneal power, mean corneal power, axis location, and astigmatic vector components (J0, J45) between devices was evaluated. Comparability was assessed using intraclass correlation, analysis of variance (ANOVA) one-way analysis, Bland-Altman analysis, and power vector analysis. Intraclass correlation was then assessed for a subset of 42 eyes with corneal astigmatism greater than or equal to 1 D.
Results:
The intraclass correlation between instruments was 0.941 for corneal power, 0.788 for corneal astigmatism, 0.932 for J0, and 0.809 for J45. The ANOVA one-way repeated P values comparing Kf, Ks, Km, corneal astigmatism, axis location, J0, and J45 were all greater than 0.05. Power vector analysis showed nearly all values between instruments to be within ±0.5 D. Finally, the intraclass correlation between instruments in eyes with ≥1 D of corneal astigmatism as measured by the T-cone was 0.946 for corneal power, 0.837 for corneal astigmatism, 0.948 for J0, and 0.950 for J45.
Conclusions:
The high level of agreement between all devices indicates the instruments may be used interchangeably. Our data suggest that this may also be true for the subset of eyes with greater than or equal to 1 D in astigmatism, highlighting the toric intraocular lens population.
More Related Videos
09:44Use of Principal Components for Scaling Up Topographic Models to Map Soil Redistribution and Soil Organic Carbon
Published on: October 16, 2018
10:59Use of Label-free Optical Biosensors to Detect Modulation of Potassium Channels by G-protein Coupled Receptors
Published on: February 10, 2014
Related Concept Videos
Topographic Surveying and Contours
Plotting of Topographic Maps
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Wave Parameters
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Evolutionary Relationships through Genome Comparisons