Analysis of Dosimetric Impacts of Cone Beam Computed Tomography-Based Volumetric Modulated Arc Therapy Planning
Manthala Padannayil Noufal1, Kallikuzhiyil Kochunny Abdullah2, Puzhakal Niyas1
1Department of Medical Physics and Radiotherapy, Baby Memorial Hospital, Calicut, India; Department of Physics, Farook College, Calicut, India; University of Calicut, Malapuram, Kerala, India.
Objective:
To quantify the Hounsfield unit (HU) variations between computed tomography (CT) and cone beam CT (CBCT) and study its impact on volumetric modulated arc therapy (VMAT) plans.
Methods:
HU number variations in CT and CBCT images were evaluated using the Catphan-504 phantom, and changes in seven different materials within the phantom (air, polymethylpentene, low-density polyethylene, polystyrene, acrylic, Delrin, and Teflon) were studied. The HU variations in half-fan and full-fan modes of CBCT were evaluated. The effect of variations in the shape of the body cross sections was assessed by reducing the body of the Catphan by 0.5 cm and 1.0 cm. CBCT-based VMAT plans in 27 patients (10 prostate, 10 brain, and 7 head and neck (HN)) were compared with corresponding CT-based plans. The dosimetric variations were assessed referring to different points on the dose volume histogram (D5%, D50%, and D95% for PTVs and D1%, Dmax, and Dmean for organs at risk). The relative percentage of difference (ΔD (%)) between CT- and CBCT-based VMAT plans were examined on these points. To evaluate the dosimetric accuracy, dose distributions were compared using Omnipro-I'mRT software. The VMAT plans were evaluated based on 3 mm-3%, 2 mm-2%, and 1 mm-1% gamma criteria.
Results:
The HU difference in CT and CBCT was highest for air, Delrin, and Teflon, whereas the difference was less than 20 HU for the other materials. The dose volume histograms of both CT- and CBCT-based plans were in excellent agreement in both phantom and patients, except in HN cases where the difference was 7%. The average 3 mm-3% gamma pass points in brain, prostate, and HN patients were 97 ± 0.2%, 96 ± 0.06%, and 93.3 ± 1.1%, respectively. The gamma pass rates reduced to 88.8 ± 0.06%, 91 ± 0.04%, and 79 ± 6% in 2 mm-2%, and further declined to 76.6 ± 0.09%, 75.2 ± 0.5%, and 60 ± 6% using the stringent 1 mm-1% gamma criteria for brain, prostate, and HN cases, respectively.
Conclusion:
Based on the results of this study, it is our belief that CBCT images can be used as a tool for evaluating the dosimetric variation in patient VMAT plans.
More Related Videos
05:49Author Spotlight: Advancing CBCT and Digital Dental Image Integration with AI-Assisted Digitization
Published on: February 23, 2024
10:23Author Spotlight: Three-Dimensional Cephalometric Landmark Annotation Demonstration on Human Cone Beam Computed Tomography Scans
Published on: September 8, 2023
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Impact Loading on a Cantilever Beam
When an object is dropped onto the free end of a cantilever, its potential energy due to gravity is...
Imaging Studies III: Computed Tomography
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Sampling Plans
Random sampling is a method where each member of the population has an equal chance of being selected for the sample. It involves selecting individuals randomly, often using random number generators or lottery-type methods. For example, when analyzing the properties of a...
Impact of Groups on Groups