Comparison of Bolus Materials to Highly Absorbent Polypropylene and Rayon Cloth
Scott Visscher1, Erin Barnett1
1Radiation Medicine Program, Bahen Chant Radiation Treatment Centre, Stronach Regional Cancer Centre, Southlake Regional Health Centre, Newmarket, Ontario, Canada.
Abstract:
This study assesses the utility of a highly absorbent polypropylene and rayon cloth (HAPRC) as a bolus material in radiation therapy. A comparison of various bolus materials used clinically at our center to the test bolus material, water-saturated HAPRC, was conducted. The reproducibility of the physical density was assessed by computerized tomography (CT) scanning the water-saturated HAPRC on a uniform density block phantom on multiple occasions, allowing the cloth to dry fully between subsequent CT scans. The cloth was found to have a consistent physical density between 0.8 and 1.0 g/cm3. To evaluate the dosimetric impact of the various bolus materials, metal oxide semiconductor field effect transistor dosimetry measurements were performed for a 6 MV treatment beam on the anterior surface of an anthropomorphic thorax phantom. The ability of the bolus material to conform to an irregular surface was assessed by CT scanning the various bolus materials overlaying the nose of an anthropomorphic head phantom and another irregularly shaped test object. The HAPRC conforms well to irregular contours and is relatively inexpensive compared to Superflab and thermoplastic sheets. Before implementing the material into the clinical setting, an appropriate process for laundering the material should be determined, for hygienic reasons.
Related Concept Videos
Two-Compartment Open Model: IV Bolus Administration
The disparity between drug input and the sum of drug transfer rates between...
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Nonlinear Pharmacokinetics: Drug Elimination for IV Bolus Injection
Following the administration of a single intravenous (IV) bolus injection, we can determine the concentration of the drug in the plasma at any given time. This calculation is achieved using a specific equation that integrates the values of Vmax and KM.
We can also...
One-Compartment Open Model for IV Bolus Administration: General Considerations
The drug's presence in the body is defined by an equation representing the difference between the rates of drug entry and exit. Key parameters—elimination rate constant,...
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
One-Compartment Open Model for IV Bolus Administration: Estimation of Clearance
In the one-compartment open model for intravenous (IV) bolus administration, clearance is estimated by dividing the elimination rate by the plasma drug concentration. This equation leverages the elimination rate constant and the apparent...