Estimation of primary radiation output for wide-beam computed tomography scanner
Atsushi Fukuda1,2, Pei-Jan P Lin1, Nao Ichikawa3,4
1Department of Radiology, Virginia Commonwealth University Medical Center, Richmond, VA, USA.
Purpose:
To estimate in-air primary radiation output in a wide-beam multidetector computed tomography (CT) scanner.
Materials And Methods:
A 6-cc ionization chamber was placed free-in-air at the isocenter, and two sheets of lead (1-mm thickness) were placed on the bottom of the gantry cover, forming apertures of 40-80 mm in increments of 8 mm. The air-kerma rate profiles were measured with and without the apertures ( , ) for 4.8 s at tube potentials of 80, 100, 120, and 135 kVp, tube current of 50 mA, and rotation time of 0.4 s. The nominal beam width was varied from 40 to 160 mm in increments of 40 mm. Upon completion of data acquisition, the were plotted as a function of the measured beam width, and the extrapolated dose rates ( ) at zero beam width were calculated by second-order least-squares estimation. Similarly, the were plotted as a function of the radiation field (measured beam width × aperture size at the isocenter), and the extrapolated dose rates ( ) were compared with the .
Results:
The means and standard errors of the with 40-, 80-, 120-, and 160-mm nominal beam widths at 120 kVp were 10.94 ± 0.01, 11.13 ± 0.01, 11.22 ± 0.01, and 11.31 ± 0.01 mGy/s, respectively, and the was reduced to 10.67 ± 0.02 mGy/s. The of 40-, 80-, 120-, and 160-mm beam widths were reduced to 10.6 ± 0.1, 10.6 ± 0.2, 10.5 ± 0.1, and 10.6 ± 0.1 mGy/s and were not significantly different from the .
Conclusions:
A method for describing the in-air primary radiation output in a wide-beam CT scanner was proposed that provides a means to characterize the scatter-to-primary ratio of the CT scanner.
More Related Videos
05:08Targeted and Selective Treatment of Pluripotent Stem Cell-derived Teratomas Using External Beam Radiation in a Small-animal Model
Published on: February 17, 2019
05:37Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Biological Effects of Radiation
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...