High-resolution lung MRI with Ultrashort-TE: 1.5 or 3 Tesla?

Guillaume Chassagnon1, Charlotte Martin2, Wadie Ben Hassen3

  • 1Radiology Department, Groupe Hospitalier Cochin-Hotel Dieu, AP-HP, Université Paris Descartes, 27 Rue du Faubourg Saint-Jacques, 75014 Paris, France; Center for Visual Computing, Ecole CentraleSupelec, 3 Rue Joliot Curie, 91190, Gif-sur-Yvette, France.

Summary

High-resolution lung MRI quality is impacted by magnetic field strength and imaging parameters. Lower image quality, including reduced subsegmental bronchus visibility and signal-to-noise ratios, was observed at 3 Tesla compared to 1.5 Tesla.

Related Concept Videos

Lung Capacity01:47

Lung Capacity

The air in the lungs is measured in volumes and capacities. Lung volume measures reflect the amount of air taken in, released, or left over after a lung function, like a single inhalation. Lung capacity measures are sums of two or more lung volume measures.
56.2K
Imaging Studies I: CT and MRI01:14

Imaging Studies I: CT and MRI

Introduction: MRI and CT scans are crucial advancements in medical imaging techniques, playing a vital role in diagnosing conditions related to the gastrointestinal (GI) system. Each scan serves distinct purposes, targets specific areas, and requires unique nursing duties.
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
847
ortho–para-Directing Activators: –CH3, –OH, –⁠NH2, –OCH301:11

orthopara-Directing Activators: –CH3, –OH, –⁠NH2, –OCH3

All ortho–para directors, excluding halogens, are activating groups. These groups donate electrons to the ring, making the ring carbons electron-rich. Consequently, the reactivity of the aromatic ring towards electrophilic substitution increases. For instance, the nitration of anisole is about 10,000 times faster than the nitration of benzene. The electron-donating effect of the methoxy group in anisole activates the ortho and para positions on the ring and stabilizes the corresponding...
7.3K
Chromatographic Resolution01:15

Chromatographic Resolution

In chromatography, a solute moves through a chromatographic column and tends to spread, forming a Gaussian-shaped band. The longer the solute spends in the column, the broader the band becomes. The broadening can lead to overlaps within the column, affecting separation effectiveness.
The effectiveness of separation can be evaluated by determining the level of separation between two neighboring peaks in a chromatogram, which represents the individual components of a sample.
In chromatography,...
2.1K
Pleura of the Lungs01:13

Pleura of the Lungs

The lungs are nestled in a cavity, shielded by the pleura. The pleura, a form of serous membrane, wraps around each lung. This membrane arrangement consists of two layers: the visceral and parietal pleurae. The visceral pleura lines the surface of the lungIn contrast, the parietal pleura is the outer layer and contacts to the thoracic wall, the mediastinum, and the diaphragm. The hilum is the point of connection between the visceral and parietal layers. The space between the parietal and...
6.9K
Radiological Investigation II: MRI and Ventilation Perfusion Scan01:30

Radiological Investigation II: MRI and Ventilation Perfusion Scan

Description
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
568