Small field correction factors determination for several active detectors using a Monte Carlo method in the Elekta
Andrea Girardi1,2, Christian Fiandra3, Francesca Romana Giglioli4
1Department of Radiotherapy, Universitair Ziekenhuis Brussel, Vrije Universiteit Brussel, Brussels, Belgium.
Abstract:
A Monte Carlo (MC) method was used to determine small field output correction factors for several active detectors (Exradin A16, Exradin A26, PTW microLion, PTW microDiamond, Exradin W1 and IBA RAZOR) for an Elekta Axesse linac equipped with circular cones. MC model of the linac was built with the GamBet software, using the Penelope code system. The dose-to-water simulation for each cone, ranging from 5 to 30 mm of diameter size, was used to calculate field factors and the results were validated together with Gafchromic EBT3 film. Output factors (OFs) were measured with the active detectors and correction factors were determined using the MC results. The MC simulations agreed with films within 1.2%. OFs measured with Exradin W1 scintillator were in agreement within 0.8% with MC simulations. The Exradin A16 and A26 microchambers under-responded for small fields relative to the MC (-13.1% and -4.6%, respectively). PTW microLion, IBA RAZOR and PTW microDiamond overestimated the output factor for the smallest field (+3.9%, +5.4 and +7.1%, respectively). The present study pointed out that it is crucial to apply the appropriate correction factors in order to provide accurate measurements in small beams geometry. The results showed that the Exradin W1 can be used for very small field dosimetry without correction factors, which shall be contrariwise employed for other detectors.
More Related Videos
08:04Cone-Enriched Cultures from the Retina of Chicken Embryos to Study Rod to Cone Cellular Interactions
Published on: March 20, 2021
07:57Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014
Related Concept Videos
Power Factor Correction
Factors Affecting Activity Coefficient
The activity coefficient value for an ion is close to one when the solution has almost zero ionic strength, i.e., when the solution shows close to ideal behavior. As the ionic strength of the solution increases from 0 to 0.1 mol/L, a...
Determining Electric Field From Electric Potential
In general, regardless of whether the electric field is uniform, it points in the direction of decreasing potential because the force on a positive...
Distance Corrections
Leveling Equipment
Transcription Factors