The Impact of Evolving Image-Guidance Processes on Initial Patient Setup for Lung Radiotherapy
Winnie Li1, Douglas J Moseley1, Tim Craig1
1Radiation Medicine Program, Princess Margaret Hospital, Toronto, Ontario, Canada; Department of Radiation Oncology, University of Toronto, Toronto, Ontario, Canada.
Purpose:
The purpose of this study was to investigate whether the implementation of three different daily image-guidance processes has altered initial (pre-imaging) patient setup practice for thoracic radiotherapy patients.
Methods:
A total of 1 997 daily cone-beam computed tomography (CBCT) images from 72 thoracic patients undergoing radical radiotherapy were retrospectively reviewed under ethics approval. Patients were analyzed in three consecutive cohorts grouped according to the image-guidance process used during radiotherapy. After initial alignment of skin marks and lasers: Process A (24 patients spanning 6 months), CBCT alignment with an action level of 3 mm, correction applied via manual couch adjustment, followed by a verification CBCT; Process B (22 patients, 5 months), CBCT alignment with an action level of 3 mm, correction applied via remote couch adjustment, followed by a verification CBCT; Process C (26 patients, 5 months), CBCT alignment with correction applied for all displacements via remote couch adjustment, with no verification scans required. Initial patient setup displacements from skin marks were determined by re-registering the initial alignment CBCT to the planning CT using automated spine matching. Patient setup displacements were compared between the three processes in the left-right (LR), cranial-caudal (CC), and anterior-posterior (AP) directions.
Results:
The mean ± 1 standard deviation of initial patient setup displacements were calculated for each cohort: Process A, 1.2 ± 2.4 mm (LR), 0.6 ± 3.5 mm (CC), and -0.8 ± 2.0 mm (AP); Process B, 0.5 ± 2.7 mm (LR), 1.2 ± 3.4 mm (CC) and -1.7 ± 2.0 mm (AP); Process C, 1.0 ± 2.5 mm (LR), 0.1 ± 3.5 mm (CC), and -2.3 ± 2.2 mm (AP). The means systematic and random uncertainties were comparable between the processes, showing similar setup error distributions.
Conclusion:
Initial skin setup practices for thoracic radiotherapy patients remain unaffected across the three image-guidance processes. Pre-imaging alignment principles and performance by radiation therapists at our center remain consistent amid technological advances.
Related Concept Videos
Initiation of Translation
First, the initiator tRNA must be selected from the pool of elongator tRNAs by eukaryotic initiation factor 2 (eIF2). The initiator tRNA (Met-tRNAi) has conserved sequence elements including modified bases at...
Initiation of Translation
Nursing Process for Patient and Caregiver Teaching II: Planning and Implementation
Nursing Process for Patient and Caregiver Teaching I: Assessment and Diagnosis
It is critical to determine the patient's learning needs during the assessment. Determination of learning needs compounds data...
Impact of Groups on Groups
Transcription Initiation
The promoters and enhancers and their accessory proteins allow tight regulation of...