Feasibility of Using Iterative Reconstruction to Reduce Radiation Dose for Computed Tomography Pulmonary Angiograms
1Mid Yorkshire Hospitals NHS Trust, Radiology Department, Pinderfields General Hospital, Wakefield, West Yorkshire, UK.
Objective:
The purpose of this study was to assess the potential for iterative reconstruction algorithms to be used to reduce patient radiation dose for computed tomography pulmonary angiography examinations while maintaining diagnostic image quality.
Materials And Methods:
This two-stage study first considered current computed tomography pulmonary angiography image quality through an audit of current practice. Secondly, an experimental design was adopted using an anthropomorphic phantom with the addition of contrast-enhanced tubes to mimic pulmonary arteries. The radiation dose was reduced by incrementally reducing the milliampere seconds value from the standard institutional protocol, and subjective and objective (contrast-to-noise ratio [CNR] and signal-to-noise ratio [SNR]) quality measures were recorded for each exposure and reconstruction.
Results:
A sample of 10 patients met the inclusion criteria for the audit, and a mean CNR of 44.2 and an SNR of 47.9 were established as baseline quality. A clear positive correlation was identified between the number of iterations and objective quality measures (CNR [r = 0.78, P < .001] and SNR [r = 0.78, P < .001]). No relationship was identified between the subjective measures and either radiation exposure (milliampere seconds) or the number of iterations. Findings suggest that a potential reduction in effective dose of 1.46 mSv (57.0%) is possible with no associated loss of image quality with the use of iterative reconstruction.
Conclusion:
Findings suggest that the potential dose reduction may be as great as 57%; however, further work is required to confirm this in a patient population.
More Related Videos
12:38Structure of HIV-1 Capsid Assemblies by Cryo-electron Microscopy and Iterative Helical Real-space Reconstruction
Published on: August 9, 2011
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Biological Effects of Radiation
Imaging Studies III: Computed Tomography
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Radiation: Applications
The average...