Background:
Cardiac surgical interventions, extracorporeal membrane oxygenation, transcutaneous coronary-artery angioplasty, and stenting are carried out while patients are being treated with the anticoagulation drug heparin. Monitoring the level and reversal of heparinization during and at the conclusion of medical and surgical procedures is a critical issue in patient care.
Methods:
We performed parallel testing of the ACCRIVA Hemochron Signature Elite ACT+ and Hemochron Response analyzer, iSTAT platform, and 2 Hepcon Hemostasis Management System (HMS) Plus analyzers for monitoring intraoperative heparin treatment. Laboratory anti-Xa assay was used as the criterion standard for heparin measurement.
Results:
Poor correlation between the 2 Hemochron analyzers was identified at 0.78. Correlation between the analyzers on the i-STAT platform was 0.97. Regression analysis revealed that i-STAT values were generally lower, by 43 seconds, than Hemochron values. The correlation between Hepcon and i-STAT activated clotting time (ACT) results was 0.94. The i-STAT ACT results were generally 23 seconds lower than the Hepcon ACT values. Correlation coefficients on comparing Hepcon ACT and i-STAT ACT using laboratory anti-Xa assay were 0.83 and 0.87, respectively. The correlation between Hepcon heparin concentration and anti-Xa results was 0.85.
Conclusions:
ACT monitoring with iSTAT offers good correlation between instruments and with the Hepcon ACT. Hepcon occupies a specific niche in cardiac operating departments because of its ability to provide additional information regarding heparin concentration; however, lack of suitable proficiency testing may impair its use. The iSTAT is a more reliable platform for broader, hospital-wide application.
Related Concept Videos
Anticoagulant Drugs: Low-Molecular-Weight Heparins
Composition of Blood Plasma
Coagulation
Coagulation
During the coagulation phase, clotting factors, or procoagulants, play a vital role in initiating and progressing the coagulation cascade. This cascade is a series of reactions...
Drug Concentrations: Measurements
Plasma...
Drug Concentration Versus Time Correlation
Two pivotal parameters are the minimum effective concentration (MEC) and the minimum toxic concentration (MTC). The MEC is the...