A Novel Application of Musculoskeletal Ultrasound Imaging
Published on: September 17, 2013
Musculoskeletal application and validation of speckle-tracking ultrasonography
Lars Henrik Frich1,2,3, Kate Lykke Lambertsen4,5,6, John Hjarbaek7
1Department of Orthopaedics and Traumatology, Odense University Hospital, J.B. Winsloewsvej 4, 5000, Odense, Denmark. lars.henrik.frich@rsyd.dk.
Background:
Diseased, injured, or dysfunctional skeletal muscles may demonstrate abnormal function and contractility. Currently, only few in vivo imaging techniques are able to characterize the contractile properties of muscle tissue. This study aimed to test the hypothesis that muscle strain can be tracked in two upper extremity skeletal muscles by speckle-tracking ultrasonography (STU) and correlates with isometric muscle contractions.
Methods:
A convenience sample of 10 healthy, adult volunteers with normal shoulder function were tested. The 5 women and 5 men had a mean age of 45 years (range: 39-59 years) and BMI < 30. STU was applied to the supraspinatus (SS) and biceps brachii (BB) muscles using a M11 L-MHz linear transducer (frequency 8-15 MHz) hooked to a Vivid E 9TM ultrasound machine. Strain validation was performed by correlating peak strain against standardized sub-maximal, isometric load conditions of the two muscles (20-80% of maximal voluntary contraction) using a custom-built muscle dynamometer based on strain-gauge technique. Data were analyzed offline using the EchoPac speckle-tracking software and were blinded to the examiner.
Results:
Intramuscular strain measured by STU in the SS and BB muscles showed moderate to strong correlations with external muscle load (SS: r = - 0.76, p < 0.0001 and BB: r = - 0.60, p < 0.0001). We found strain to vary from approximately 10-20% during increasing submaximal, isometric conditions.
Conclusions:
We demonstrate that STU can be applied on healthy skeletal musculature (SS and BB muscles). The observed correlations between strain and isometric contractions suggest a valid technique. However, the concept of measuring muscle strain non-invasively needs further investigation for validity, accuracy, responsiveness, and reliability before its therapeutic and research potential can be realized.
More Related Videos
09:05Transthoracic Speckle Tracking Echocardiography for the Quantitative Assessment of Left Ventricular Myocardial Deformation
Published on: October 20, 2016
10:25Technical Approach for Infrared Tracking for Soft Tissue Navigation with a Holographic Head-Mounted Display and Preclinical Validation
Published on: September 2, 2025
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Reliability and Validity
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...
Imaging Studies II: Ultrasonography
Data Validation
Nursing assessment guides are generally based on holistic models rather than medical...
Data Validation
Key parameters for method validation include: