Slow injector-to-column sample transport to maximize resolution in liquid chromatography: Theory versus practice
Fabrice Gritti1, Nobuo Tanaka2
1Waters Corporation, Instrument/Core Research/Fundamental, Milford, MA 01757, USA.
Abstract:
A simple chromatographic model including extra-column sample bandspreading is built and validated experimentally. It is used to quantify the advantage of slowly transporting under isocratic elution of the injected sample band along the pre-column volume of the chromatographic system compared to the conventional injection method at constant flow rate. For fast analyses (<30 s), the model predicts a maximum relative gain of 20% for the isocratic peak capacity of weakly retained compounds (k < 1) at the price of a 60% increase in retention time (+20 s) for short (3.0 cm long) narrow-bore (2.1 mm i.d.) columns packed with sub-2 μm particles and for low-dispersion (∼2 μL2 extra-column volume variance) vHPLC systems. These predictions were confirmed experimentally using a 2.1 mm × 30 mm column packed with 1.9 μm XBridge-C18 for the rapid separation of small molecules in 25 s at 40 °C using a mixture of methanol and water (75/25, v/v) as the eluent. For longer analyses (>30 s), the model allows for the determination of the optimum and slow injection speed that maximizes the trade-off between the analysis time and the isocratic peak capacity. For 2.1 mm × 20 mm columns packed with 2 μm particles and a highest retention factor of 10, the optimum injection speed is about 10-15% of the operating column flow rate for a maximum relative gain of 5% in overall isocratic peak capacity at constant elution time. Overall, the proposed injection method is directly applicable for enhanced resolution of weakly retained polar compounds in RPLC without the need of changing separation conditions, which may strongly increase the retention times of the late apolar eluters.
More Related Videos
11:00Untargeted Metabolomics from Biological Sources Using Ultraperformance Liquid Chromatography-High Resolution Mass Spectrometry UPLC-HRMS
Published on: May 20, 2013
10:21Automated Hydrophobic Interaction Chromatography Column Selection for Use in Protein Purification
Published on: September 21, 2011
Related Concept Videos
Types Of Column Chromatography
Gel Filtration Chromatography
When the...
Diffusion on Chromatography Columns
Longitudinal diffusion occurs when the solute molecules in the mobile phase diffuse from the more concentrated center of the chromatographic band to the more dilute regions on either side, both towards and against the flow direction. This...
Principles Of Column Chromatography
Column Efficiency: Plate Theory
A higher number of theoretical plates signifies better column efficiency and improved separation capabilities. Plate height affects bandwidth and separation quality; it is inversely...
Column Efficiency: Rate Theory
During elution, a solute molecule experiences numerous transitions between stationary and mobile phases, exhibiting irregular residence times in...
Silica Gel Column Chromatography: Overview
Polar components tend to bind strongly to the silica gel, causing them to move slowly through the column. In contrast, nonpolar compounds...