Use of a Rat Model to Study Ventral Abdominal Hernia Repair
Published on: October 2, 2017
Increasing hernia size requires higher GRIP values for a biomechanically stable ventral hernia repair
F Kallinowski1, D Gutjahr1, M Vollmer2
1Klinik für Allgemein-, Viszeral- und Transplantationschirurgie, Chirurgische Universitätsklinik Heidelberg, Im Neuenheimer Feld 110, D-69120, Heidelberg, Germany.
Background:
Increasing hernia sizes lead to higher recurrence rates after ventral hernia repair. A better grip might reduce the failure rates.
Material And Methods:
A biomechanical model delivering dynamic intermittent strain (DIS) was used to assess grip values at various hernia orifices. The model consists of a water-filled aluminium cylinder covered with tissues derived from pig bellies which are punched with a central defect varying in diameter. DIS was applied mimicking coughs lasting for up to 2 s with peak pressures between 180 and 220 mmHg and a plateau phase of 0.1 s. Ventral hernia repair was simulated with hernia meshes in the sublay position secured by tacks, glue or sutures as needed to achieve certain grip values. Grip was calculated taking into account the mesh: defect area ratio and the fixation strength. Data were assessed using non-parametric statistics.
Results:
Using a mesh classified as highly stable upon DIS testing (DIS class A) a reduced overlap without fixation led to early slippage (p < 0.001). With the application of 16 fixation points, transmural sutures were better than tacks with Securestrap® being better than Absorbatack® (p < 0.001). Plotting the likelihood of a durable repair as a function of the calculated grip higher grip values were needed with increasing hernia diameter to achieve biomechanical stability. This is important for clinical work since the calculated grip values both from a registry and from published data tend to drop as hernia sizes increase indicating biomechanical instability.
Conclusion:
The experimental work reported here demonstrates for the first time that higher grip values should be reached when repairing larger ventral hernias.
Related Concept Videos
Biodiversity and Human Values
Professional Values
The values that are the foundation of the nursing profession are altruism, autonomy, human dignity, and social justice.
First, altruism refers to the concern for the welfare and well-being of others without personal...
Overview of DNA Repair
Chemically...
Critical Values
Mismatch Repair
z Scores and Unusual Values
This score indicates how far a value is from the mean in terms of standard deviation. For example, if a data value has a z score of +1, the researcher can infer that the particular data value is one standard deviation above the mean. If another data...