In Vitro Thrombosis Test for Ventricular Assist Devices
Published on: March 21, 2025
The design modification of advanced ventricular assist device to enhance pulse augmentation and regurgitant flow
Takuma Miyamoto1, Nicole Byram1, Jamshid H Karimov1
1Department of Biomedical Engineering, Lerner Research Institute, Cleveland Clinic, Cleveland, Ohio.
Abstract:
The new Advanced ventricular assist device (Advanced VAD) has many features such as improving pulsatility and preventing regurgitant flow during pump stoppage. The purpose of this study was to evaluate the effects of design modifications of the Advanced VAD on these features in vitro. Bench testing of four versions of the Advanced VAD was performed on a static or pulsatile mock loop with a pneumatic device. After pump performance was evaluated, each pump was run at 3000 rpm to evaluate pulse augmentation, then was stopped to assess regurgitant flow through the pump. There was no significant difference in pump performance between the pump models. The average pulse pressure in the pulsatile mock loop was 23.0, 34.0, 39.3, 33.8, and 37.3 mm Hg without pump, with AV010, AV020 3S, AV020 6S, and AV020 RC, respectively. The pulse augmentation factor was 48%, 71%, 47%, and 62% with AV010, AV020 3S, AV020 6S, and AV020 RC, respectively. In the pump stop test, regurgitant flow was -0.60 ± 0.70, -0.13 ± 0.57, -0.14 ± 0.09, and -0.18 ± 0.06 L/min in AV010, AV020 3S, AV020 6S, and AV020 RC, respectively. In conclusion, by modifying the design of the Advanced VAD, we successfully showed the improved pulsatility augmentation and regurgitant flow shut-off features.
Related Concept Videos
Design Example: Flow Through a Fire Extinguisher
The key to understanding how the...
Design Example: Flow of Oil Through Circular Pipes
Histone Modification
Acetylation
The enzyme histone acetyltransferase adds acetyl group to the histones. Another enzyme, histone...
Pulse
The pulse serves as a clinical...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls...
Spreading of Chromatin Modifications
Writers
The writer...