Related Experiment Video
Updated: Jan 25, 2026

Bergmeyer Glucose Quantification for Microbiological Samples
Published on: January 17, 2025
Use of the Gravimetric Diluter in Microbiological Work
1Department of Animal Sciences and Industry, Kansas State, University, Manhattan, Kansas 66506-1600.
Abstract:
Manual dilutions of food samples (e.g., 1:10 dilution) for microbiological investigations are cumbersome. We tested an automatic diluter (Model GD-600 Gravimetric Diluter, Spiral System, Inc., Bethesda, MD) for accuracy and sterility for microbiological work and found that the system provided ±0.1 to +0.2 g of accuracy in delivering a commanded amount of diluent. The delivered diluent was essentially sterile, with a few samples giving 1 to 4 colonies per ml after plating. With this system, weighing and dilution are fast, easy, and accurate.
More Related Videos
Related Concept Videos
Solution Concentration and Dilution
History of Microbiology
Three-Dimensional Microscopy in Microbiology
Two-Dimensional Microscopy in Microbiology
Key Techniques in Microbiology
Formation of Dilute Urine
Filtrate Osmolarity in the PCT
Initially, as the filtrate passes through the proximal convoluted tubule (PCT), its...