Methodology paper: a novel phantom setup for commissioning of scanned ion beam delivery and TPS
O Jäkel1,2,3, B Ackermann4, S Ecker4
1Heidelberg Ion-Beam Therapy Center (HIT) at the University Hospital, Heidelberg, Germany. o.jaekel@dkfz.de.
Background:
Commissioning of treatment planning systems (TPS) and beam delivery for scanned light ion beams is an important quality assurance task. This requires measurement of large sets of high quality dosimetric data in anthropomorphic phantoms to benchmark the TPS and dose delivery under realistic conditions.
Method:
A novel measurement setup is described, which allows for an efficient collection of a large set of accurate dose data in complex phantom geometries. This setup allows dose measurements based on a set of 24 small volume ionization chambers calibrated in dose to water and mounted in a holder, which can be freely positioned in a water phantom with various phantoms mounted in front of the water tank. The phantoms can be scanned in a CT and a CT-based treatment planning can be performed for a direct benchmark of the dose calculation algorithm in various situations.
Results:
The system has been used for acceptance testing in scanned light ion beam therapy at Heidelberg Ion Beam Therapy Center for scanned proton and carbon ion beams. It demonstrated to be useful to collect large amounts of high quality data for comparison with the TPS calculation using various phantom geometries.
Conclusion:
The setup is an efficient tool for commissioning and verification of treatment planning systems. It is especially suited for dynamic beam delivery, as many data points can be obtained during a single plan delivery, but can be adapted also for other dynamic therapies, like rotational IMRT.
More Related Videos
11:03Nanoscale Characterization of Liquid-Solid Interfaces by Coupling Cryo-Focused Ion Beam Milling with Scanning Electron Microscopy and Spectroscopy
Published on: July 14, 2022
10:58Focused Ion Beam Fabrication of LiPON-based Solid-state Lithium-ion Nanobatteries for In Situ Testing
Published on: March 7, 2018
Related Concept Videos
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Common Ion Effect
Precipitation of Ions
The equation that describes the equilibrium between solid calcium carbonate and its solvated ions is:
Ion Channels
Ion channels are specialized integral membrane proteins on the plasma membrane that allow...
Leaky Scanning
Formation of Complex Ions