Screening for hemoglobin C using Technicon H-6000: the effect of formalin
Acta Haematologica
|January 1, 1987
Summary
Formalin interferes with red blood cell lysis, causing the automated H-6000 cell counter to fail when processing blood samples with hemoglobin C. This study identifies formalin as the cause of this critical analytical error.
Area of Science:
- Hematology
- Clinical Chemistry
- Laboratory Science
Background:
- Automated cell counters are essential for accurate blood analysis.
- The H-6000 cell counter has demonstrated an inability to process blood samples containing hemoglobin C.
- Previous studies have not identified the specific cause of this analytical failure.
Purpose of the Study:
- To investigate the cause of red blood cell lysis failure in the H-6000 cell counter for samples with hemoglobin C.
- To identify the interfering substance responsible for the H-6000's performance issue.
Main Methods:
- Comparative analysis of blood samples with and without hemoglobin C.
- Investigation of potential interfering agents, including common laboratory fixatives.
- Assessment of red blood cell lysis efficiency using the H-6000 cell counter.
Main Results:
- Formalin was identified as the substance inhibiting red blood cell lysis.
- The presence of formalin directly correlates with the failure of the H-6000 cell counter to lyse red blood cells in samples with hemoglobin C.
- Hemoglobin C samples without formalin exposure were processed correctly.
Conclusions:
- Formalin is the causative agent for the failure of red blood cell lysis in the H-6000 cell counter when analyzing samples with hemoglobin C.
- Awareness of formalin's interference is crucial for accurate hematological analysis.
- Recommendations for sample handling and instrument maintenance should be considered to mitigate this issue.