Concurrent validity and reliability of a low-cost gait analysis system for assessment of spatiotemporal gait
Maijke van Bloemendaal1, Anita Beelen, Rob F M Kleissen
1Rehabilitation, Merem, 1213 VX Hilversum, The Netherlands. mbloemendaal@merem.nl.
Objective:
To determine the concurrent validity and reliability of a low-cost spatiotemporal gait analysis system for clinical use in rehabilitation medicine.
Design:
Cross-sectional study.
Subjects:
Thirty-three healthy adults.
Methods:
The spatiotemporal gait analysis system consists of a video camera placed perpendicular to a 10-m walkway and calibrated for spatial reference. The conditions evaluated in this study were: barefoot walking at comfortable and slow speed, toe and shod walking using a stationary camera setup and barefoot walking at comfortable speed using a moving camera setup. The GAITRite® was used as reference.
Results:
High intraclass correlation coefficients (ICC≥ 0.97; 95% lower limit confidence intervals (95% CIs) ≥ 0.77) were found between systems for step and stride length, and step, stance and stride time, across setups and conditions. Standard error of measurement and Bland-Altman repeatability coefficients were ≤ 2.4% and ≤ 6.3%, respectively. A minimum of 4 footsteps was required to obtain ICC >0.90 and coefficient of variation < 10%. For double support and swing time, ICCs were generally low (ICC≥ 0.21). Inter-rater reliability was excellent for step length, step and stance time (ICC≥ 0.94; lower limit 95% CIs ≥ 0.86).
Conclusion:
The spatiotemporal gait analysis system is valid and reliable for assessing spatiotemporal parameters in different walking conditions. However, the validity of double support and swing time could not be confirmed.
More Related Videos
06:25Paw-Print Analysis of Contrast-Enhanced Recordings PrAnCER: A Low-Cost, Open-Access Automated Gait Analysis System for Assessing Motor Deficits
Published on: August 12, 2019
06:54Clinical-oriented Three-dimensional Gait Analysis Method for Evaluating Gait Disorder
Published on: March 4, 2018
Related Concept Videos
Reliability and Validity
Noncompartmental Analysis: Miscellaneous Pharmacokinetic Parameters
One key aspect of the noncompartmental approach is determining a drug's total clearance. This can be done by dividing the drug dose by the area under the concentration-time curve from zero to infinity. The area under the concentration-time curve represents the drug's...
Wave Parameters
Distribution Reliability and Automation
Data Validation
Nursing assessment guides are generally based on holistic models rather than medical...
Data Validation
Key parameters for method validation include: