Application of a standard autoclaving protocol does not harm structural integrity of two-piece zirconia abutments
Dieter Pils1,2, Ralf J Baeppler2, Rüdiger Junker2
1Pils Zahntechnik GmbH, Bruck an der Glocknerstrasse, Austria.
Objective:
Aim was to assess the influence of a standard autoclaving protocol on the retention of zirconia abutments glued on titanium bases in two-piece implant abutments.
Materials And Methods:
Twenty-four CAD/CAM-generated zirconia abutments were adhesively cemented on prefabricated titanium bases. Before mechanical and thermal aging, the specimens were divided into two groups. Group 1 was subjected to a standard steam autoclaving protocol and Group 2 remained untreated (control). The tensile strength in all specimens was evaluated by a standardized pull-off test limited to a maximum force of 1000 N.
Results:
Eleven samples in both groups could be subjected to pull-off testing. Ten samples in Group 1 and three samples in Group 2 failed, while all others reached the maximum pull-off force. This difference was statistically significant. The mean retention values for the failed samples were 694.53 ± 369.10 N in Group 1 and 890.78 ± 25.90 N in Group 2. This difference was not statistically significant.
Conclusions:
A standard autoclaving protocol does not reduce detachment force of two-piece zirconia abutments.
Clinical Relevance:
Clinical sterilization processes as recommended by regulatory authorities seem to be harmless to the structural integrity of two-piece zirconia implant abutments, at least with regard to the retention of the components.
More Related Videos
Related Concept Videos
Application of Integration: Problem Solving
Testing a Claim about Standard Deviation
The hypothesis testing for the claim of population standard deviation (or variance) requires the data and samples to be random and unbiased. The population distribution also must be normal. There is no specific requirement on the sample size as the estimation is based on the chi-square distribution.
As a first step, the hypothesis (null and alternative) concerning the claim about...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Intermolecular Forces
Intermolecular Forces in Solutions
When the strengths of the intermolecular forces of attraction between solute and solvent species in a solution are no different than those present in the separated components, the solution is formed with no accompanying energy change. Such a solution is called an ideal solution. A mixture of ideal gases (or gases such as helium and argon,...
Electromotive Force