Retrospective fractional dose reduction in Tc-99m cardiac perfusion SPECT/CT patients: A human and model observer
P Hendrik Pretorius1, Albert Juan Ramon2, Michael A King3
1Department of Radiology, University of Massachusetts Medical School, 55 Lake Avenue North, Worcester, MA, 01655, USA. Hendrik.Pretorius@umassmed.edu.
Background:
In the ongoing efforts to reduce cardiac perfusion dose (injected radioactivity) for conventional SPECT/CT systems, we performed a human observer study to confirm our clinical model observer findings that iterative reconstruction employing OSEM (ordered-subset expectation-maximization) at 25% of the full dose (quarter-dose) has a similar performance for detection of hybrid cardiac perfusion defects as FBP at full dose.
Methods:
One hundred and sixty-six patients, who underwent routine rest-stress Tc-99m sestamibi cardiac perfusion SPECT/CT imaging and clinically read as normally perfused, were included in the study. Ground truth was established by the normal read and the insertion of hybrid defects. In addition to the reconstruction of the 25% of full-dose data using OSEM with attenuation (AC), scatter (SC), and spatial resolution correction (RC), FBP and OSEM (with AC, SC, and RC) both at full dose (100%) were done. Both human observer and clinical model observer confidence scores were obtained to generate receiver operating characteristics (ROC) curves in a task-based image quality assessment.
Results:
Average human observer AUC (area under the ROC curve) values of 0.725, 0.876, and 0.890 were obtained for FBP at full dose, OSEM at 25% of full dose, and OSEM at full dose, respectively. Both OSEM strategies were significantly better than FBP with P values of 0.003 and 0.01 respectively, while no significant difference was recorded between OSEM methods (P = 0.48). The clinical model observer results were 0.791, 0.822, and 0.879, respectively, for the same patient cases and processing strategies used in the human observer study.
Conclusions:
Cardiac perfusion SPECT/CT using OSEM reconstruction at 25% of full dose has AUCs larger than FBP and closer to those of full-dose OSEM when read by human observers, potentially replacing the higher dose studies during clinical reading.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
07:54Implantation and Monitoring by PET/CT of an Orthotopic Model of Human Pleural Mesothelioma in Athymic Mice
Published on: December 21, 2019
Related Concept Videos
Drug Dosing: Geriatric Patients
Drug Dosing: Obese Patients
Observational Studies
There are three types of observational studies – Prospective, retrospective, and cross-sectional.
Prospective Study
Prospective studies, also known as longitudinal or cohort studies, are carried out by collecting future data from groups sharing similar characteristics. One...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Bioavailability Study Design: Single Versus Multiple Dose Studies