Related Experiment Video
Updated: Jan 25, 2026

Minimally Invasive Thumb-sized Pterional Craniotomy for Surgical Clip Ligation of Unruptured Anterior Circulation Aneurysms
Published on: August 11, 2015
Single-center experience with the new-generation Derivo embolization device for ruptured and unruptured intracranial
Marius G Kaschner1, Athanasios Petridis2, Bernd Turowski3
1Department of Diagnostic and Interventional Radiology, Medical Faculty, University of Duesseldorf, Duesseldorf, Germany - marius.kaschner@med.uni-duesseldorf.de.
Background:
The aim of this study is to evaluate safety and efficacy of the new-generation Derivo embolization device in complex ruptured (RIA) and unruptured (UIA) intracranial aneurysms.
Methods:
Retrospective analysis of 32 patients with complex RIAs and UIAs treated with Derivo at our hospital from November 2015 to December 2018. Clinical safety was defined as absence of death, transient attack, absence of minor and major stroke, and Derivo associated hemorrhage. Treatment efficacy was assessed angiographically (DSA) immediately after treatment and at 6-month follow-up according to the O'Kelly-Marotta (OKM) Grading Scale (from A=total filling to D=no filling; prolongation of stasis 1=arterial to 3=venous phase).
Results:
Thirty-two patients with 39 aneurysms were treated with 42 Derivos. In five aneurysms, additional coiling was performed. Deployment was technically successful in all cases. Two patients developed a procedure related minor stroke (one transient). In one patient bleeding due to an inflammatory aneurysmatic wall process occurred 20 days after retreatment and in one patient a stroke due to in-stent thrombosis occurred when dual platelet inhibition (PI) was switched to permanent single PI 12 month after FD treatment. No treatment related deaths were observed. Initial DSA revealed three OKM D, six OKM C, five OKM B, and 25 OKM A. Six-month follow-up for DSA and clinical evaluation was available in 20/32 patients (62.5%), 26 of 39 aneurysms (66.7%) and revealed 73.1% complete and 3.8% subtotal occlusion (OKM D: 19 of 26, OKM C3: one of 26).
Conclusions:
Treatment of complex RIAs and UIAs with the new-generation Derivo appeared to be safe and effective in this single-center case series for ruptured and unruptured intracranial aneurysms. Immediate DSA revealed a significant flow modulation; and 6-month follow-up showed a high occlusion rate.
Related Concept Videos
Lattice Centering and Coordination Number
Types of Unit Cells
Imagine taking a large number of identical...
Pulmonary Embolism I: Introduction
Center of Gravity
Center of Gravity
To determine its location, the principle of moments can be utilized by dividing the object into...
Center of Mass
The knowledge of the center of mass can also help us to describe and predict the motion of objects. For example, when a ball is thrown...
Aneurysm I: Introduction