Reliability and Validity of A-Mode Ultrasound to Quantify Body Composition
Nathan Hendrickson1, John Davison, Luke Schiller
1University of Iowa Hospitals and Clinics, Iowa City, IA.
Objective:
Immobilization and inadequate nutrition after high-energy trauma result in loss of lean muscle mass. Body composition [percent body fat (%BF), fat mass (FM), fat-free mass (FFM)] is traditionally quantified with techniques that require expensive equipment and/or ionizing radiation. The purpose of this study was to assess reliability and validity of amplitude-mode (A-mode) ultrasound (US) to quantify body composition for use in a randomized clinical trial assessing interventions to prevent muscle loss in young individuals after high-energy musculoskeletal trauma.
Methods:
Thirty-one volunteer subjects underwent body composition assessments using A-mode US and air displacement plethysmography (ADP). Independent raters performed 2 serial US measures. The same raters performed body composition assessment in 12 consecutive subjects that sustained acute musculoskeletal trauma indicated for surgery. Test-retest and inter-rater reliability were assessed using intraclass correlation coefficient (ICC). Agreement between US and ADP was assessed with Bland-Altman analysis.
Results:
Test-retest and inter-rater reliability was excellent for volunteer subjects, with ICC values (%BF, FM, and FFM) of 0.87, 0.90, and 0.99 (rater 1); 0.80, 0.82, and 0.98 (rater 2). FFM measured by US was strongly correlated with ADP measures (r = 0.9635, P < 0.05). Bland-Altman analysis demonstrated no systematic bias between US and ADP measures of body composition. In trauma subjects, the inter-rater reliability was excellent, with ICCs of 0.96, 0.98, and 0.99 for %BF, FM, and FFM, respectively.
Conclusions:
A-mode US is a relatively low-cost tool that provides reliable estimates of body composition and is a viable alternative for monitoring body composition in young, healthy patients.
More Related Videos
09:21Muscle Function Obtained with Motion Mode Ultrasound and Surface Electromyography during Core Endurance Exercise
Published on: August 25, 2022
06:48Author Spotlight: Advancements in 3D Optical Imaging for Comprehensive Body Composition Assessment in Modern Research
Published on: June 7, 2024
Related Concept Videos
Reliability and Validity
Composite Bodies
Composite bodies have widespread applications in mechanical engineering, from automobiles to aircraft to rockets. For example, an automobile wheel comprises...
Composition of Body Fluids
What is a Mode?
There can be more than one mode in a data set if multiple values have the same highest frequency. For instance, suppose that the Statistics exam scores of 20 students are: 50; 53; 59; 59; 63; 63; 72; 72; 72; 72; 72; 76; 78; 81; 83; 84; 84; 84; 90; 93. Here, the mode is 72, as it occurs most frequently, five times.
A data set with two modes is called bimodal. For example,...
Classifying Matter by Composition
According to its composition, the matter can be classified into two broad categories — pure substances and mixtures.
A pure substance is a form of matter that has a constant composition throughout with uniform properties. For example, any sample of sucrose has the same composition and same physical properties, such as melting point, color, and sweetness, regardless of the source from which it is isolated.
A mixture is composed of two or...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...