The Effect of a Bone Groove Against Cochlear Implant Electrode Migration
Martin Leinung1, Silke Helbig, Youssef Adel
1Department of Otolaryngology, Head and Neck Surgery, University Hospital Frankfurt, Frankfurt am Main, Germany.
Hypothesis:
The risk of electrode migration in cochlear implant (CI) surgery can be reduced by securing the electrode lead in a bone groove in the chorda-facial angle.
Background:
Electrode migration is an important complication in CI surgery, which affects hearing performance, may induce pain and facial nerve stimulation, and requires surgical revision. A potential method to secure the electrode is to use a bone groove created in the facial recess to fixate the electrode lead. This surgical measure is a standard procedure for lateral wall electrodes (LWE) in our clinic since 2013.
Methods:
Retrospective analysis of consecutive CI cases of a tertiary referral center from 2006 to 2016 and comparison of incidence rates between group A (without bone groove, years 2006-2012) and group B (with bone groove fixation, years 2013-2016). Clinical cases were reviewed with respect to electrode type, migration length, time interval to migration, and findings during revision surgery.
Results:
Seventeen cases of electrode migration were found in a total of 1,603 cochlear implantations. Only LWE designs were affected by migration. The cumulative incidence for LWE in group B (0.5%) was significantly lower than in group A (3.7%, p < 0.01). The two migration cases in group B revealed insufficient geometric design of the bone grooves. Only one patient experienced a remigration after revision surgery.
Conclusion:
The fixation in a bone groove in the chorda-facial angle is an effective measure against migration of LWE. Proper surgical execution is mandatory to ensure a tight and enduring fit.
Related Concept Videos
Standard Electrode Potentials
Migration
Electrodes: Overview
There are two main types of electrodes in electrochemical cells. The first type, known as the working or indicator electrode, has a potential that is sensitive to the analyte's concentration and reacts to changes in...
Cell Migration
Cell Migration
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...