Chanan Shaul1, Christofer W Bester1, Stefan Weder1
1Department of Surgery - Otolaryngology, University of Melbourne, Royal Victorian Eye and Ear Hospital, East Melbourne, Australia.
Objectives/Hypothesis:
Spikes in cochlear implant impedance are associated with inner ear pathology after implantation. Here, we correlate these spikes with episodes of hearing loss and/or vertigo, with a comparison between lateral wall and peri-modiolar electrode arrays.
Methods:
Seven hundred seventy recipients of Cochlear's slim-straight, lateral wall electrode (CI422), or peri-modiolar (CI512) electrode were investigated for impedance spikes. Impedance fluctuations were defined as a median rise of ≥ 4 kΩ across all intracochlear electrodes from baseline measurements taken 2 weeks after switch-on. Medical records were analyzed from 189 of the 770 patients.
Results:
The slim straight, lateral wall electrode was found to spike in impedance at a significantly higher rate than the peri-modiolar array (17% vs 12%). The peri-modiolar electrode tended to spike in impedance earlier than the slim-straight electrode. Impedance spikes were found to significantly correlate with medical events (hearing loss, vertigo, or tinnitus). Overall, in the "spike" group, 42 of 75 patients (56%) demonstrated a clinical event during the impedance spike, whereas 26 of 114 patients (22%) of the "non-spike" group had a clinical event. This significant difference existed with both implant types.
Conclusion:
These results demonstrate a small, but significant increase in impedance spikes in lateral wall electrodes, and support the relationship between spikes in cochlear implant impedances and postoperative inner-ear events, including the loss of residual hearing and vertigo. Monitoring cochlear implant impedance may be a method for early detection, and so the prevention, of these events in the future.
Related Concept Videos
Plant Cell Wall
Anatomy of the Ear
Impedances and Admittance
Impedance is a measure of resistance to sinusoidal current flow in an AC circuit. Unlike their behavior in DC circuits, where inductors appear as short circuits and capacitors as open circuits, the behavior of these components in AC...
Impedance Combination
Lateralization
Series Impedances: Three-Phase Line
Using Kirchhoff's laws, an integro-differential equation for the network is derived. This equation accounts for unbalanced phase currents, which may induce return currents through neutral wires and the earth, seeking the least impedance path. Earth return conductors can replace the...