Using Q Suture to Enhance Resistance to Gap Formation and Tensile Strength of Repaired Flexor Tendons
Published on: June 3, 2020
Rotator cuff tendon tissue cut-through comparison between 2 high-tensile strength sutures
Brett D Owens1, Joseph Algeri2, Vivian Liang2
1Department of Orthopaedic Surgery, Warren Alpert School of Medicine, Brown University, Providence, RI, USA.
Dynacord suture caused less tendon cut-through than FiberWire in rotator cuff repairs. This finding is crucial for preventing mechanical failure in tendon surgeries.
Area of Science:
- Orthopedic Surgery
- Biomaterials Science
- Musculoskeletal Research
Background:
- High-tensile strength sutures can cause mechanical failure by cutting through tendon tissue during rotator cuff repairs.
- Suture-induced tendon damage, or 'cheese-wiring,' is a significant concern in orthopedic surgery.
- Evaluating novel suture materials is essential for improving surgical outcomes.
Purpose of the Study:
- To compare the tendon-cutting potential of Dynacord suture versus FiberWire suture.
- To assess suture performance in a controlled laboratory setting using ovine rotator cuff models.
- To determine if Dynacord or FiberWire causes less tissue damage during cyclic loading.
Main Methods:
- Seven pairs of ovine shoulders were used, with Dynacord and FiberWire sutures (USP size No. 2) passed through infraspinatus tendons.
- Samples underwent 1000 cycles of cyclic testing to simulate surgical stress.
- The amount of tendon cut-through was measured, with >5 mm defined as clinical failure.
Main Results:
- Dynacord sutures resulted in significantly less tendon cut-through (2.69 ± 1.02 mm) compared to FiberWire sutures (3.72 ± 1.14 mm) (P = .012).
- In matched-pair analysis, FiberWire showed greater cut-through in 13 of 14 specimens.
- Two FiberWire specimens exhibited complete tendon cut-through (>5 mm), indicating clinical failure.
Conclusions:
- Dynacord suture demonstrated superior performance by causing less tendon cut-through than FiberWire in an ovine rotator cuff model.
- The findings suggest Dynacord may reduce the risk of mechanical failure associated with suture pull-through.
- Further clinical studies are warranted to evaluate patient outcomes and healing rates with Dynacord suture.
Related Concept Videos
Tensile Strength Considerations of Concrete
The dimensions and shape of a concrete specimen...
Relation Between Tensile Strength and Compressive Strength of Concrete
Acid Strength and Molecular Structure
In the absence of any leveling effect, the acid strength of binary compounds of hydrogen with nonmetals (A) increases as the H-A bond strength decreases down a group in the periodic table. For group 17, the order of increasing acidity is HF < HCl < HBr < HI. Likewise, for group 16, the order of increasing acid strength is H2O < H2S < H2Se < H2Te. Across a row in the periodic table, the acid strength of binary hydrogen compounds increases with increasing...
Strength of Cement
For compressive strength tests, ASTM C 109-05 standards prescribe a cement-sand mix ratio of 1:2.75 and a water/cement ratio of 0.485 for making 2-inch cubes. These cubes are mixed, cast, and cured in saturated lime water at 23°C until testing. Flexural strength testing, outlined in...
Sutures of the Skull
Sutures are immobile joints between adjacent bones of the skull. The narrow gap between the bones is filled with dense, fibrous connective tissue that unites the bones. The long sutures located between the skull bones are not straight but instead follow irregular, tightly twisting paths. These twisting lines tightly...
The Equilibrium Binding Constant and Binding Strength