Jamie Toms1,2, Sheyne Martin2, Adam P Sima3
1Department of Neurosurgery, Virginia Commonwealth University, Richmond, Virginia, USA.
Background:
Frameless stereotactic surgery utilizing fiducial-based (FB) registration is an established tool in the armamentarium of deep brain stimulation (DBS) surgeons. Fiducial-less (FL) registration via intraoperative CT, such as the O-arm, has been routinely used in spine surgery, but its accuracy for DBS surgery has not been studied in a clinical setting.
Objective:
We undertook a study to analyze the accuracy of the FL technique in DBS surgery and compare it to the FB method.
Methods:
In this prospective cohort study, 97 patients underwent DBS surgery using the NexFrame and the O-arm registration stereotactic system. Patients underwent FB (n = 50) registration from 2015 to 2016 and FL (n = 47) O-arm registration from 2016 to 2017.
Results:
The radial errors (RE) and vector/euclidean errors of FB and FL registration were not significantly different. There was no difference in additional passes between methods, but there was an increase in the number of RE ≥2.5 mm in the FL method.
Conclusion:
Although there was no statistically significant difference in RE or the need for additional passes, the increased number of errors ≥2.5 mm with the FL method (17 vs. 4% in FB) indicates the need for further study. We concluded that O-arm images of the implants should be utilized to assess and correct for this error.
Related Concept Videos
Muscles that Move the Arm
The pectoralis major has two origins. Its clavicular head originates on the medial half of the clavicle. In contrast, the sternocostal head originates on the costal cartilages of ribs 1-6, the sternum, and the aponeurosis of the external oblique of the...
Comparing Copy Number Variations and SNPs
Copy number variations or CNVs are the structural variations that cover more than 1kb of DNA sequence. The single nucleotide polymorphism (SNP), on the other hand, is a single nucleotide change or a point mutation that is found in more than 1%...
Comparing the Survival Analysis of Two or More Groups
Comparing Experimental Results: Student's t-Test
Comparative Excretory Systems
Comparing Mitochondrial, Chloroplast, and Prokaryotic Genomes