Inadvertent Removal of a Right Ventricular Pacemaker Lead by a Knotted Transvenous Pacing Wire
Evan J Wiens1, Colette M Seifer2, Clarence Khoo2
1Department of Internal Medicine, Max Ready College of Medicine, University of Manitoba, Winnipeg MB, Canada.
Abstract:
We report an unusual complication of permanent pacemaker (PPM) implantation in a patient with a temporary transvenous pacemaker (TVP) in situ, in which the TVP lead formed a knot around the newly-placed right ventricular (RV) lead, complicating TVP removal. The case underscores the fact that suboptimal TVP lead placement, such as looping in the RV, can result in complications during implantation of a permanent pacemaker.
How To Cite This Article:
Wiens EJ, Seifer CM et al. Inadvertent Removal of a Right Ventricular Pacemaker Lead by a Knotted Transvenous Pacing Wire. Indian J of Crit Care Med 2019;23(2):102-103.
More Related Videos
08:05Optimization of Transesophageal Atrial Pacing to Assess Atrial Fibrillation Susceptibility in Mice
Published on: June 29, 2022
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Related Concept Videos
Household Wiring And Electrical Safety
Magnetic Field Due To A Thin Straight Wire
Magnetic Field Due to Two Straight Wires
Magnetic Force On Current-Carrying Wires: Example
Phase-lead and Phase-lag Controllers
Time and frequency -Domain Interpretation of Phase-lead Control
The design of phase-lead control involves the strategic placement of poles and zeros to balance steady-state error and system...