Development and Assessment of a Microcomputed Tomography Compatible Five Degrees-of-Freedom Knee Joint Motion
Alexandra M Blokker1, Alan M Getgood2, Nathan J Curiale3
1Department of Biomedical Engineering,Robarts Research Institute,Western University,London, ON N6A 5B9, Canada.
Abstract:
Currently available knee joint kinematic tracking systems fail to nondestructively capture the subtle variation in joint and soft tissue kinematics that occur in native, injured, and reconstructed joint states. Microcomputed tomography (CT) imaging has the potential as a noninvasive, high-resolution kinematic tracking system, but no dynamic simulators exist to take advantage of this. The purpose of this work was to develop and assess a novel micro-CT compatible knee joint simulator to quantify the knee joint's kinematic and kinetic response to clinically (e.g., pivot shift test) and functionally (e.g., gait) relevant loading. The simulator applies closed-loop, load control over four degrees-of-freedom (DOF) (internal/external rotation, varus/valgus rotation, anterior/posterior translation, and compression/distraction), and static control over a fifth degree-of-freedom (flexion/extension). Simulator accuracy (e.g., load error) and repeatability (e.g., coefficient of variation) were assessed with a cylindrical rubber tubing structure and a human cadaveric knee joint by applying clinically and functionally relevant loads along all active axes. Micro-CT images acquired of the joint at a loaded state were then used to calculate joint kinematics. The simulator loaded both the rubber tubing and the cadaveric specimen to within 0.1% of the load target, with an intertrial coefficient of variation below 0.1% for all clinically relevant loading protocols. The resultant kinematics calculated from the acquired images agreed with previously published values, and produced errors of 1.66 mm, 0.90 mm, 4.41 deg, and 1.60 deg with respect to anterior translation, compression, internal rotation, and valgus rotation, respectively. All images were free of artifacts and showed knee joint displacements in response to clinically and functionally loading with isotropic CT image voxel spacing of 0.15 mm. The results of this study demonstrate that the joint-motion simulator is capable of applying accurate, clinically and functionally relevant loads to cadaveric knee joints, concurrent with micro-CT imaging. Nondestructive tracking of bony landmarks allows for the precise calculation of joint kinematics with less error than traditional optical tracking systems.
Related Concept Videos
One-Degree-of-Freedom System
A one-degree-of-freedom system is defined by an independent variable that determines its state and behavior. One example of a one-degree-of-freedom system is a simple harmonic oscillator, such as a...
Degrees of Freedom
For example, suppose there are three unknown numbers whose mean is 10; although we can freely assign values to the first and second numbers, the value of the last number can not be arbitrarily assigned.
Degrees of Freedom
For example, suppose there are three unknown numbers whose mean is 10; although we can freely assign values to the first and second numbers, the value of the last number can not be arbitrarily...
Knee Joint
A total of seven ligaments support the knee joint. The patellar ligament, which is also attached to the quadriceps femoris...
Development of the Limb Synovial Joints
The mesenchymal stem cells differentiate into chondrocytes that form the hyaline cartilage, and later the cartilaginous model of the bone. This model further transforms into a bone. This process is known as endochondral ossification.
During development, the limbs...
Degree of Unsaturation
The degree of unsaturation for hydrocarbons is U = (2C + 2 − H) / 2, where C is the number of carbon atoms and H is the number of hydrogen atoms.