Accuracy of Doppler blood pressure measurement in continuous-flow left ventricular assist device patients

Song Li1, Jennifer A Beckman1, Nathan G Welch2

  • 1Division of Cardiology, Department of Medicine, University of Washington, Seattle, WA, 98195, USA.

ESC Heart Failure
|May 18, 2019
PubMed

Insights

Accurate blood pressure (BP) measurement is crucial for continuous-flow ventricular assist device (CF-VAD) patients. Doppler opening pressure offers a reliable non-invasive method for BP assessment in these individuals.

Area of Science:

  • Cardiovascular Medicine
  • Biomedical Engineering

Background:

  • Accurate blood pressure (BP) monitoring is critical for managing patients with continuous-flow ventricular assist devices (CF-VADs) to mitigate risks such as stroke.
  • Current non-invasive BP measurement methods may lack accuracy in CF-VAD populations.

Purpose of the Study:

  • To evaluate the accuracy of the Doppler opening pressure method for BP measurement in CF-VAD patients.
  • To compare Doppler opening pressure accuracy against the gold standard invasive arterial line method.

Main Methods:

  • A longitudinal cohort study involving HeartMate II and HVAD patients.
  • Simultaneous measurement of arterial line BP and Doppler opening pressure.
  • Analysis of correlation and agreement between methods, including the impact of arterial pulsatility.

Main Results:

  • 1933 paired readings from 154 CF-VAD patients were analyzed.
  • Doppler opening pressure showed good correlation with invasive mean arterial pressure (r=0.742, P<0.0001).
  • The method more accurately approximated mean arterial pressure (mean error 2.4 mmHg) than systolic BP (mean error -8.4 mmHg), with no significant effect from arterial pulsatility.

Conclusions:

  • Doppler opening pressure demonstrates good accuracy for BP measurement in CF-VAD patients.
  • This non-invasive technique is a viable alternative to invasive monitoring.
  • Doppler opening pressure is recommended as the standard non-invasive BP measurement in CF-VAD patients.
Abstract

Related Concept Videos

Assessing Blood pressure using a doppler ultrasound01:19

Assessing Blood pressure using a doppler ultrasound

To obtain accurate blood pressure measurements in clinical settings, especially when traditional methods are insufficient, healthcare professionals utilize the Doppler ultrasound technique. This method uses high-frequency sound waves to detect blood flow within the arteries, which is crucial for patients with conditions that complicate circulatory system assessment.
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
2.5K
Measurement of Blood Pressure01:17

Measurement of Blood Pressure

Assessing blood pressure is a standard procedure executed in virtually all medical environments. The method utilized today was established over a hundred years ago by an innovative Russian doctor, Dr. Nikolai Korotkoff. The soft ticking noise, known as Korotkoff sounds, heard while taking blood pressure readings results from turbulent blood flow within the vessels. The apparatus required for this procedure includes a sphygmomanometer, a blood pressure cuff attached to a gauge, and a...
2.7K
Uncertainty in Measurement: Accuracy and Precision03:37

Uncertainty in Measurement: Accuracy and Precision

Scientists typically make repeated measurements of a quantity to ensure the quality of their findings and to evaluate both the precision and the accuracy of their results. Measurements are said to be precise if they yield very similar results when repeated in the same manner. A measurement is considered accurate if it yields a result that is very close to the true or the accepted value. Precise values agree with each other; accurate values agree with a true value. 
100.5K
Equipments Used To Measure Blood Pressure01:30

Equipments Used To Measure Blood Pressure

Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
3.1K